User Manual
42 GMC-I Gossen-Metrawatt GmbH
V/Hz, Ω, Temperature,   and A/Hz Measurements
5.7.1 Direct and Pulsating Current Measurement, Direct Connection, 
A DC and A (DC+AC) 
(METRAHIT EXTRA, METRAHIT ETECH and 
METRAHIT E
SPECIAL only)
➭ First disconnect supply power from the measuring circuit or 
the power consumer (1), and discharge any capacitors.
➭ In accordance with the current to be measured, turn the rotary 
switch to A   or A
  .
➭ Select the current type appropriate for the measured quantity 
by briefly pressing the FUNC | ENTER multifunction key. Each 
time the key is pressed, the instrument is switched back and 
forth between A DC and A (DC + AC)
TRMS
, which is indicated 
by means of an acoustic signal. The current type is indicated at 
the LCD by means of the DC or the (DC+AC)
TRMS
 symbol.
➭ Safely connect the measuring instrument (without contact 
resistance) in series to the power consumer (2) as shown.
➭ Switch supply power to the measuring circuit back on (3).
➭ Read the display. Make a note of the measured value if the 
instrument is not being operated in the memory mode or the
transmission mode.
➭ Disconnect supply power from the measuring circuit or the 
power consumer (1) once again, and discharge any 
capacitors.
➭ Remove the test probes from the measuring point and return 
the measuring circuit to its normal condition.
COM mA A
~
R
x
~
R
x
1
3
2
I > 10 A
!
A
A
003
.
50
A
DC
003
.
50
A
DC
AC
TRMS
A
A
METRAHIT EXTRA
METRAHIT ESPECIAL
6 A / 10 A (16 A max. 30 s)
600 μA / 6 mA
60 mA / 600 mA
6 A / 10 A (16 A max. 30 s)
8
.
8
.
8
.
8
.
8
!
I > 16 A
Strom
Measuring Ranges:
Current measurement 
may only be performed 
with installed batteries!
FUNC
ENTER
METRAHIT ETECH
60 mA / 600 mA
6 A / 10 A (16 A max. 30 s)