V/Hz, Ω, Temperature,   and A/Hz Measurements
Alternating Voltage Level Measurement (dB)
Voltage level measurement is used in order 
to ascertain overall attenuation or boosting 
of a transmission system (shown here as a 
4-pole setup). 
Voltage level [dB]
where U
1
 = U
REF
 (reference level)
Result > 1: boosting
Result < 1: attenuation
➭ Manually select the measuring range for the voltage amplitude. 
When the instrument is switched to dB measurement, the pre-
viously selected voltage measuring range remains active.
➭ Repeatedly press the FUNC | ENTER multifunction key until unit 
of measure dB appears at the display. 
Lowest measurable frequencies and maximum permissible 
voltages are included in chapter 8, “Technical Data”.
The level measurement function is now activated. The measured 
value is calculated based upon the RMS value of the alternating 
voltage component relative to the measuring range (600 mV 
…1000 V), and displayed.
The default setting for th reference level is 0 dB = 0,775 V (1 mW 
to 600 Ω). This value can be adjusted in the „SET “ menu (see also 
chapter 6.4): 
 1nFo 
 ...  SET   rAtE  ...  dbrEF
 0 ..001 ... 99 V 
   
Note
No terminal resistors have been integrated into the device. It 
performs measurement with a high input impedance of 
9MΩ. 
Input impedance for voltage measurement is listed under 
technical data. 
In order to be able to perform correct measurement at non-
terminated devices under test, the terminating resistor must 
be connected to the terminals. Be sure to take power loss at 
the terminating resistor into consideration!