Control Functions
4.4 Measured Value Storage: DATA (auto-hold / compare)
An individual measured value can be automatically “frozen” with 
the DATA function (auto-hold). This is useful, for example, when 
contacting the measuring points with the test probes requires 
your full attention. After the measuring signal has been applied 
and the measured value has settled in in accordance with the 
“condition” listed in the table below, the measured value is frozen 
at the digital display and an acoustic signal is generated. The test 
probes can now be removed from the measuring points, and the 
measured value can be read from the digital display. If the 
measuring signal falls below the value specified in the table, the 
function is reactivated for storage of the next value.
Measured Value Comparison (DATA Compare)
If the currently frozen value deviates from the first saved value by 
less than 100 digits, the acoustic signal is generated twice. If 
deviation is greater than 100 digits, only a brief acoustic signal is 
generated.
Note
The DATA function has no effect on the analog display, at 
which the current measured value continues to appear. 
However, when the digital display if “frozen”, the decimal 
point is fixed as well (fixed measuring range, symbol: MAN). 
The selected measuring range should not be manually 
changed as long as the DATA function is active.
The DATA function is deactivated by pressing and holding the 
DATA/MIN/MAX key (for approx. 1 second), when the measuring 
function is changed or when the instrument is switched off and 
back on again.
1)
Reactivation results from falling short of specified measured value limits.
2)
Two acoustic signals are generated the first time a measured value is 
saved as a reference value. For subsequent data hold, two acoustic 
signals are only generated if the currently frozen value deviates from the 
first saved value by less than 100 digits.
Key: MV = measured value, R = measuring range
DATA 
Function
Press 
DATA /
Min/Max
Condition Response from Instrument
Measuring 
Function
Measuring 
Signal
Display
Acous-
tic
MV
Digital
DATA
Activate Brief blinks Once
Save
(stabilized 
measured 
value)
V, A, Hz, dB, 
F, M H z ,  %
> 10% 
of R
Is dis-
played
Static
Once
Twice 
2)
Ω   
≠ 0L
Reactivate 
1)
V, A, Hz, dB, 
F, M H z ,  %
< 10%  
of R 
Stored 
MV
Blinks
Ω 
= 0L
Change to
Min/Max
Brief See table in chapter 4.4.1
Exit Long
Is cleared Is cleared
Twice
V, A, Hz
t [s]
60,000
5,000
Activated
Reactivated
Save
Save
of measuring range
dB, F, MHz, %
d
d