Control Functions
4 Control Functions
4.1 Selecting Measuring Functions and Measuring Ranges
The rotary switch is linked to the automatic socket blocking 
mechanism, which only allows access to two connector jacks for 
each function. Be certain to remove the appropriate plug from its 
respective jack before switching to and from the “A” functions. 
The socket blocking mechanism prevents the user from inadvert-
ently turning the selector switch to impermissible functions after 
the measurement cables have been plugged in to the instrument.
4.1.1 Automatic Range Selection
The multimeter is equipped with auto-ranging for all measuring 
functions, except for temperature measurement, diode and 
continuity testing, and the MHz measuring function. Auto-ranging 
is active as soon as the instrument is switched on. The instrument 
automatically selects the measuring range which allows for 
highest possible resolution of the applied quantity. When the 
instrument is switched to frequency measurement, the previously 
selected voltage measuring range remains active.
The Auto-Ranging Function
The multimeter is switched automatically to the next higher range 
at ±(60000 d + 1 d → 06000 d), and to the next lower range at 
±(05000 d - 1 d → 50000 d). 
Exception, capacitance measurement:
The multimeter is switched automatically to the next higher range 
at ±(6000 d + 1 d → 0 600 d), and to the next lower range at 
±(0500 d - 1 d → 5000 d). 
4.1.2 Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges can be 
selected manually in accordance with the following table by 
pressing the MAN / AUTO button.
The desired measuring range can then be selected with the 
 or 
 scroll key.
The instrument is automatically returned to range selection when 
the MAN / AUTO key is pressed, the rotary switch is activated or the 
instrument is switched off and back on again.
Overview: Auto-Ranging and Manual Range Selection 
* Via manual measuring range selection only
Function Display
MAN / AUTO
Manual mode active: 
utilized measuring range is fixed
MAN
 or 
Range switching sequence for: 
V:
600 mV*
 ↔
 6 V
 ↔
 60 V
 ↔
 600 V
 ↔
 1000 V 
Hz(V AC):
600 Hz
 ↔
 6 kHz
 ↔
 60 kHz
 ↔
 300 kHz
MHz
600 Hz
 ↔
 6 kHz
 ↔
 60 kHz
 ↔
 600 kHz 
↔
 1 MHz
Ω
:600 
Ω ↔ 
6 k
Ω ↔
 60 k
Ω ↔ 
600 k
Ω ↔
 6 M
Ω ↔
 60 M
Ω
A: METRAHIT E
XTRA
: 600
 μ
A
 ↔
 6 mA
 ↔
 60 mA
↔
 600 mA
 ↔
 6 A
 ↔
 10 A (16 A) 
METRAHIT E
TECH
: 
60 mA
 ↔
 600 mA
 ↔
 6 A
 ↔
 10 A (16 A) 
METRAHIT E
SPECIAL
: 6 A 
↔
 10 A (16 A)
Hz (A AC):
 600 Hz
 ↔
 6 kHz
 ↔
 60 kHz
A:  A
(mV) 600 mV 
↔
 6 V
A
(mA) 60 mA 
↔
 600 mA
 ↔
 6 A
siehe chapter 5.7.3 ff.
F: METRAHIT E
XTRA
 und 
METRAHIT E
TECH
: 
60 nF
 ↔
 600 nF
 ↔
 6 
μ
F
 ↔
 60
 μ
F
 ↔
 600
 μ
F
MAN
MAN / AUTO
Return to automatic measuring range selection —