User Manual

GMC-I Gossen-Metrawatt GmbH 41
V/Hz, Ω, Temperature, and A/Hz Measurements
5.7 Current Measurement
Notes Regarding Current Measurement
The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents are otherwise not
indicated, and the instrument may be damaged.
The METRAHIT ESPECIAL multimeter does not have a fuse for the cur-
rent measuring range. It’s designed for measurements in current
transformer circuits and is approved for measuring category 600 V
CAT II
.
Set up the measuring circuit in a mechanically secure fashion, and
secure it against inadvertent breaks. Select conductor cross-sec-
tions and lay out connections such that they do not overheat.
An intermittent acoustic signal warns of current greater than 10 A.
An continuous acoustic signal warns of current greater than 16 A.
The input for the current measuring range is equipped with a fuse
link. Maximum permissible voltage for the measuring
circuit (= rated voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only! The fuse must have a
breaking
capacity of at least 30 kA
.
If the fuse for the active current measuring range blows, “
FUSE
appears at the digital display, and an acoustic signal is
generated at the same time.
If a fuse should blow, eliminate the cause of overload before plac-
ing the instrument back into service!
Fuse replacement is described in chapter 9.3.
Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are in-
cluded in chapter 8, “Technical Data”, in the table entitled “Meas-
uring Functions and Measuring Ranges” in the “Overload
Capacity” column.
Scope of Functions, Current Measurement, Direct Connection
* ) Special model without fuse for measuring-current transformer
Scope of Functions, Current Measurement via Current Clamp Sensors
Scope of Functions for Current Measurement with Current Clamps Transformer
Function METRAHIT
E
XTRA
METRAHIT
E
TECH
METRAHIT
E
SPECIAL
METRAHIT
E
BASE
A AC / Hz ~ 600 μA
6/60/600 mA
6 A / 10 (16) A
60/600 mA
6 A / 10 (16) A
6 A / 10 (16) A
A AC+DC TRMS
600 μA
6/60/600 mA
6 A / 10 (16) A
60/600 mA
6 A / 10 (16) A
6 A / 10 (16) A
A DC 600 μA
6/60/600 mA
6 A / 10 (16) A
60/600 mA
6A / 10 (16) A
6 A / 10 (16) A
1000 V fuse
••* )
Function METRAHIT
EXTRA
METRAHIT
ETECH
METRAHIT
ESPECIAL
METRAHIT
EBASE
Transformation
Factor
••••
A AC / Hz
••••
A AC+DC
••••
A DC
••••
Hz (A AC)
... 60 kHz ... 60 kHz ... 60 kHz ... 60 kHz
Function METRAHIT EXTRA METRAHIT ETECH METRAHIT ESPECIAL
Transformation Factor
••
A AC / Hz
••
Hz (A AC)
... 60 kHz ... 60 kHz ... 60 kHz