User Manual

GMC-I Gossen-Metrawatt GmbH 23
Control Functions
4.5 Measurement Data Recording (METRAHIT EXTRA only)
The METRAHIT E
XTRA is capable of recording measurement data
using an adjustable sampling rate for long periods of time in the
form of measurement series. Data are stored to a battery backed
memory module, and are retained even after the multimeter is
switched off. The system acquires measured values relative to
real-time.
Stored measured values can subsequently be read out with the
help of METRAwin 10 software. The only prerequisite is a PC which
is connected by means of an interface cable to the
USB X-TRA bi-directional interface adapter, which is in turn
plugged onto the METRAHIT E
XTRA. See also chapter 7, “Interface
Operation”.
Memory Parameters Overview (METRAHIT EXTRA only)
The STORE Menu Function
First set the sampling rate for memory mode operation (see rAtE
parameter in chapter 6.4) , and then start memory mode oper-
ation.
First select the desired measuring function and an appropriate
measuring range.
Check the battery charge level before starting long-term
measurement recordings (see chapter 6.3).
Connect the NA X-TRA power pack if required.
Starting Recording via Menu Functions
Switch to the “SET ” mode by pressing MEASURE | SETUP and se-
lect the “StorE ” menu.
1nFo
... StorE 000.0% StArt
Memory mode operation is started by pressing FUNC | ENTER.
STORE appears underneath the analog display and indicates
that memory mode operation has been activated.
StoP ” appears at the digital display.
Press MEASURE | SETUP in order to return to the measuring
function.
Parameter Page: header
CLEAr
24: Clear Memory
EMpty
24: Clear Memory – appears after CLEAr
0CCvP
24: Querying Memory Occupancy
rAtE
50: rAtE – set the sampling rate (METRAHIT EXTRA only)
StArt
23: Starting Recording via Menu Functions
StoP
24: Ending Recording
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER