User Manual

84 GMC-I Messtechnik GmbH
Technical Data
Acoustic Signals
For voltage Intermittent signal at above 600 V
For current Intermittent signal at above 10 A
Continuous signal at above 16 A
Fuse
Fuse link FF (UR) 10 A/1000 V AC/DC;
10 mm x 38 mm; breaking capacity of
30 kA at 1000 V AC/DC;
protects the current measurement input in
the 600 μA to 10 A ranges
Electrical Safety
Per IEC 61010-1:2001/VDE 0411-1:2002
Protection class
II
Measuring category CAT III CAT IV
Operating voltage 600 V 300 V
Pollution degree
2
Test voltage
5.2 kV~
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1: 2006 class B
Interference immunity EN 61326-1: 2006
EN 61326-2-1: 2006
Ambient Conditions
Accuracy range 0 °C to +40 °C
Op. temp. range
T
A
10 °C to +50 °C *
Storage temp. range 25 °C to +70 °C (without batteries)
Relative humidity 40 to 75%, no condensation allowed
Elevation To 2000 m
Deployment
Indoors, except within specified ambient
conditions
* Exception: measurement of current > 10 to 16 A, operation at up to 40 °C
Mechanical Design
Housing Impact resistant plastic (ABS)
Dimensions 200 x 87 x 45 mm
(without rubber holster)
Weight Approx. 0.4 kg with batteries
Protection Housing: IP 52
(pressure equalization via housing)
Table Excerpt Regarding Significance of
IP-Codes
IP XY
(1
st
digit X)
Protection Against Foreign
Object Entry
IP XY
(2
nd
digit Y)
Protection Against
Penetration by Water
5 Dust protected 2 Dripping (at angle of 15°)