User Manual
GMC-I Messtechnik GmbH 65
Device and Measuring Parameters
bLiGht – Disable Deactivation of Background Illumination
If necessary, automatic deactivation of background illumination
can be disabled with this parameter setting or via the interface.
1nFo
... SET system
system t iME
... system blight
blight set: apoff / on
Note
Each time the instrument is switched on, the bLiGht
parameter is reset to bLiGht = APoFF!
0.diSP – Show/Hide Leading Zeros
This parameter determines whether or not leading zeros will
appear in the measured value display.
1nFo
... SET system
system t iME
... system 0.diSP 0.diSP set:
00000.0 : with leading zeros (default value)
0.0 : leading zeros suppressed
6.4.2 EVEntS Submenu
EVEntS – Event Counter Sampling Rate and Trigger Thresholds
An event is recorded whenever at least one measured value has
fallen short of the L-trig threshold, and subsequently at least one
measured value has exceeded the H-trig threshold. Voltage signals
with a repetition frequency of less than 500 Hz, or less than 2 Hz,
are recorded (events rate of 0.001 or 0.5 seconds).
1nFo ... SET system ... events
events trig/events rate DC
h-trig set : -60000 ... +60000
l-trig set : -60000 ... +60000
events rate : 0 ..001 /0..5 s