User Manual

48 GMC-I Messtechnik GmbH
Measurements: V/Hz/dB – PQ/ – MHz/%Ω/nS/R
SL
– /
Temp – /m – W/Wh – A/Hz/thd
Use current clamp transformers
only. Power measurement is not
possible with current clamp sen-
sors equipped with voltage out-
put.
Note:
See also section 4.5.2, "Power
and Energy Measurement in the
Memory Mode”.
Min-Max Function in Power Measurement
Instantaneous power is displayed with the switch in the W setting.
The Min-Max function is activated with the DATA/MIN/MAX key. The
minimum and maximum values for active, reactive an apparent
power are display as of the beginning of the power measurement,
along with time of occurrence (date and time). Previous Min-Max
values can be cleared without exiting the function by pressing the
ZERO key.
This function differs from the general Min-Max function insofar as
measurement is performed continuously in the background, even if
the Min-Max display is not shown, and while the instrument is in
the energy measuring mode.
Measurement of Active, Apparent and Reactive Power, and Power Factor – W, VA, VAr, PF
COM
V
Tem p
Ω
AmA
N
L
~
W
000
.
00
W
DC
AC
TRMS
W
000
.
0
Bandwidth:
... 1 kHz
Wh
mV
000
.
0
mA
VA
000
.
00
VA
DC
AC
TRMS
pf
:
0
.
00
VAr
μ
μ
000
.
00
DC
AC
TRMS
000
.
0
mV
000
.
0
mA
VAr
μ
MAN / AUTO
DATA/MIN/MAX
MAN / AUTO
DATA/MIN/MAX
MAN / AUTO
DATA/MIN/MAX
* With current clamp trans. only
3.6 mW / 36 mW / 360 mW
3.6 W / 36 W / 360 W
3.6 kW / 6 kW / 36 kW* /
W measuring ranges:
360 kW* / 3.6 MW*
> 30 V AC or > 35 V DC:
Warnings regarding dangerous
voltage:
PF measuring range:
–(0 ... 0.99): capacitive
+(0 ... 0.99): inductive
±1:
no phase shifting
10 A (max. 5 min.)
16 A (max. 30 s)
Overload capacity at 600 V: