Datasheet

METRAHITENERGY
TRMS SYSTEM Multimeter
6
GMC-I Messtechnik GmbH
Influencing Quantities and Influence Error
1
The 60 mV DC range and thermocouple measurement are sensitive to temperature
fluctuation: For this reason, specified values are not valid 30 until minutes after
ambient temperature has stabilized.
2
Power limiting: frequency x voltage max. 6 x 10
6
V x Hz for U > 100 V
3
The accuracy specification for frequency response is valid as of a display value of
10% of the measuring range for both measuring modes with the TRMS converter
in the A AC and A (AC+DC) ranges.
4
Frequency response up to 100 kHz, > 60 kHz plus 5%
Response Time (after manual range selection)
Data Interface
Type
Optical via infrared light through the housing
Data transmission Serial, bidirectional (not IrDa compatible)
Protocol Device specific
Baud rate 38,400 baud
Functions Select/query measuring functions
and parameters
Query momentary measurement data
– Read out stored measurement data
The USB X-TRA plug-in interface adapter (see accessories) is
used for adaptation to the PC’s USB port.
Internal Measured Value Storage
Memory capacity 16 MBit for approx. 300,000 measured
values with indication of date and time
Power Supply
Battery 2 each 1.5 V AA batteries, alkaline
manganese per IEC LR6 (2 each 1.2 V
NiMH rechargeable batteries also possible)
Service life With alkaline manganese: approx. 120 hrs.
Battery indicator Battery capacity display with battery sym-
bol in 4 segments: .
Querying of momentary battery voltage via
menu function.
Power OFF function The multimeter is switched off
automatically
:
– When battery voltage drops below
approx. 1.8 V
– If none of the keys or the rotary switch
are activated for an adjustable duration
of 10 to 59 min. and the multimeter is
not in the continuous operation mode
(assuming the instrument is not set to
power measurement or mains analysis)
Power pack socket
If the NA X-TRA power pack has been
plugged into the instrument, the batteries
are disconnected automatically.
Rechargeable batteries can only be
recharged externally.
Power pack voltage: 5.1 V 0.2 V
Influencing
Quantity
Sphere of
Influence
Measured Quantity /
Measuring Range
Influence Error per 10 K
(...% rdg. + ... d)
Temperature
0 C ... +21 C
and
+25 °C... +40 C
60 mV
1
0.2 + 5
600 mV ... 600 V 0.1 + 5
600 mV 0.3 + 20
V , 6 ... 600 V 0.2 + 10
600 ... 60 M, nS 0.2 + 5
A , , 0.2 + 10
60 nF ... 6 F, km 1 + 5
60, 600 F3 + 5
Hz, dB 0.2 +10
Diode measurement 0.3 + 5
RSL measurement 1 + 10
Pt100 / Pt1000 0.5 + 10
Type K thermocouple
1
0.2 + 10
Power measurement: V 0.3 + 10
Power measurement: A 0.2 + 5
W, VA, Wh, VAh 0.5 + 10
Influenc-
ing Qty.
Meas. Quantity /
Measuring Range
Sphere of Influence
Intrinsic Uncertainty
3
 ... % rdg. + ... d)
Fre-
quency
V
AC
600.00 mV
> 15 Hz ... 45 Hz 3 + 30
> 65 Hz ... 1 kHz 2 + 30
> 1 kHz ... 20 kHz 3 + 30
> 20 kHz ... 100 kHz
4
3.5 + 30
4
6.0000 V
...
60.00 V
> 15 Hz ... 45 Hz 2 + 30
> 65 Hz ... 1 kHz 1 + 30
> 1 kHz ... 20 kHz 3 + 30
> 20 kHz ... 100 kHz
4
3.5 + 30
4
600.00 V
2
> 15 Hz ... 45 Hz 2 + 30
> 65 Hz ... 1 kHz 1 + 30
> 1 kHz ... 20 kHz 3 + 30
A
AC
600.00 A
...
10.0000 A
> 15 Hz ... 45 Hz
3 + 25
> 65 Hz ... 10 kHz
Influencing
Quantity
Sphere of
Influence
Measured Quantity Influence Error
Relative
humidity
75%
3 days
instrument off
V A, , F, Hz, dB, C 1 x intrinsic uncertainty
Battery voltage 1.8 ... 3.2 V V A, , F, Hz, dB, C
Included in
intrinsic uncertainty
Influencing
Quantity
Sphere of Influence
Measured Quantity /
Measuring Range
Damping
Common mode
interference
voltage
Interference quantity max. 600 V V > 120 dB
Interference quantity max. 600 V
50 Hz ... 60 Hz, sine
6 V , 60 V > 80 dB
600 V > 70 dB
Series-mode
interference
voltage
Interference quantity V ,
respective nominal value of the
measuring range,
max. 600 V , 50 Hz ... 60 Hz, sine
V> 50dB
Interference quantity max. 600 V V > 100 dB
Measured Quantity/
Measuring Range
Digital Display
Response Time
Measured Quantity
Jump Function
V, V, dB
AV , A
1.5 s
From 0 to 80%
of upper range limit value
600 ... 6 M 3s
From to 50%
of upper range limit value
nS, R
SL
3s
60 M 8s
Continuity
(acoustic signal)
<50ms
C (Pt100) Max. 3 s
1.5 s
60 nF ... 600 FMax. 2s
From 0 to 50%
of upper range limit value
>10 Hz 1.5 s