METRAHITCAL
Calibrator
Internal clock
Time format DD.MM.YYYY hh:mm:ss
Resolution 0.1 s
Accuracy 1 minute per month
Temp. influence 50 ppm/K
Display
LCD panel (65 x 36 mm) with digital display including simulation
unit of measure and various special functions
Background Illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Display / char. height 7-segment characters
Main display: 1 x 6 digits, 12 mm
Auxiliary displays: 2 x 6 digits, 7 mm
Max. resolution 30000
Polarity display “–” (minus sign) is displayed
Refresh Rate 2 times per second, every 500 ms
Power Supply
Battery 2 AA size batteries,
alkaline manganese per IEC LR6
(2 ea. 1.2 V NiMH rechargeable battery
also possible)
Service life With alkaline manganese (2600 mAh)
If voltage drops below 1.8 V, the
instrument is switched off automatically.
Battery Indicator Battery capacity display with battery
symbol in 4 segments: .
Querying of momentary battery voltage via
menu function.
Power Saving Circuit
The device is switched off automatically if none of the controls are
activated for a period of approximately 10 minutes. The simulator
is switched off after a period of only 5 minutes (sockets are cur-
rent and voltage-free). Automatic shutdown can be deactivated.
Power pack socket
If the NA X-TRA power pack is plugged in,
the installed batteries are disconnected
automatically.
Rechargeable batteries can only be
recharged externally.
Fuses
Fuse link FF0,63A/700V, 6,3 mm x 32 mm,
min. 1,5 kA switching capacity
Electrical Safety
Protection class II per
EN 61010-1:2001/VDE 0411-1:2002
Operating voltage Max. 50 V
Measuring category I (250 V)
Pollution degree 2
Test voltage 500 V~ per EN 61010-1:2001/VDE 0411-
1:2002
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1: 2006 class B
Interference immunity EN 61326-1: 2006
EN 61326-2-1: 2006
Data Interface
Typ e
Optical via infrared light through the housing
Data transmission Serial, bidirectional (not IrDa compatible)
Protocol Device specific
Baud rate 38,400 baud
Functions Set/query cal. functions and parameters
The USB X-
TRA plug-in interface adapter (see accessories) is used
for adaptation to the PC’s USB port.
Ambient Conditions
Accuracy range 0 C +40 C
Operating temp. range 10 C +50 C
Storage temp. range 25 C +70 C (without batteries)
Relative humidity 40% 75%,
no condensation allowed
Elevation To 2000 m
Mechanical Design
Housing Impact resistant plastic (ABS)
Dimensions 200 x 87 x 45 mm
(without protective rubber holster)
Weight Approx. 0.35 kg with batteries
Protection IP 54
(pressure equalization via the housing)
Table Excerpt Regarding Significance of IP
Codes
Calibration Function Power
Consump-
tion
Service life
mV, thermocouple 55 mA 45 h
15 V 240 mA 10 h
, RTD 85 mA 30 h
Sink, 20 mA 310 mA 8 h
Source, 20 mA 310 mA 8 h
IP XY
(1
st
char. X)
Protection against pene-
tration by solid particles
IP XY
(2
nd
char. Y)
Protection against
penetration by water
5 Dust protected 4 Splashing water