Enter or select:
How the measured value will be entered to the program:
Multimeter Input
The measured value is
the help of an adapter.
Manual Input
The displayed actual value for the device to be calibrated is read and entered
using the PC’s keyboard.
Variable Output
The calibration signal is continuously changed at the monitor or the calibrator
by means of manual tracking with a decade device until the display at the
measuring or recording instrument coincides with the specified setpoint.
When the measured value read in via the port is valid (Wait)
Which measured value is expected from the device under test (UUT
Expected Value) and how accurate the value should be (UUT Specification)
Whether or not an auto-message will appear at the monitor screen
Whether or not an individual message will be read out to the user before
executing a procedure step
se, or is created with the help of
for all of the calibration points until the procedure has been
tions in any given procedure