User manual
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 17
#regval Measured value
#fregval Measured value with measuring function
#rdev Measured difference (absolute value)
#frdev Measured difference with measuring function
#rdig Measured difference in digits
#rprc
Measured difference as percentage of the specification
#result
Assessment: pass/limit/fail/---
#text
Comment regarding the test step
#udig
Measuring uncertainty in digits
#uval
Measuring uncertainty (absolute value)
#urat
Measuring uncertainty ratio
General Parameters
Variable Name
Parameter
#passprc
Pass percentage for limit value
#measured
Number of measured steps
#passed
Number of passed steps
#limit
Number of limit steps (between limit value and failure)
#fail
Number of failed steps
►
Adapt structure and column content
in accordance with your needs and save the
edited file as a Word 97 or 2003 template (*.DOT), or as Word template with
macros (*.DOTM) with a new name.
Running the Word macro and transferring the calibration data t
o the Word report template
►
Start Microsoft
®
Word and open the desired Word template with macro.
►
Start the included AutoNew.MAIN macro by clicking Macros and then View
Macros in the View menu.
⇒
Respond to the prompt asking whether or not you want to receive data from the
calibration program with [Yes]:
⇒
In the window which then appears, you’re prompted to open a test procedure
file. If this file has already been opened, click [Cancel].
⇒
Now you’re prompted to start calibration (if you haven’t already done so) and
then to select the export test data menu command.
⇒
The data are transferred to the Word file and completion is acknowledged with a
ready message.
Calibration data can be transferred to Excel in a similar way. However, the macro and
the report templates are located in separate files in the p
rogram directory in this case.
►
Start Microsoft
®
Excel and open the KLT2DOC.XLM Excel macro file.
⇒
Two icons now appear underneath the add-ins with the designations KLT:Get
calibrator data and KLT:Close.
►
Click the KLT:Get calibrator data icon and select the report template file
(KLT2XLSx.XLT) to be opened in the window which then appears.