User manual
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 16
template. Corresponding report templates with integrated macros can be found in
the program directory to this end.
Editing the Microsoft
®
Word report template included with the program
►
Start Microsoft
®
Word
and open the Word template with macro (KLT2DOC.DOT)
located in the program directory.
The included macro (AutoNew.MAIN) can import the following parameters from the
open test procedure with results
based on the variable names:
Report Header Parameter
s
Variable Name
Parameter
#lab
Issuing laboratory
#calby
Calibrated by
#caldate
Calibration date
#nextdate
Next calibration date (= calibration date + 12 months)
#uutmake
Manufacturer of the device under test
#uuttype
Type designation of the device under test
#uutnumb
Serial number of the device under test
#uutdsc
Object (type of device under test)
#envtemp
Ambient temperature
#envhmdt
Relative humidity
#meatype
Type of measurement
#certnum
Certificate number
#caldev
Type designation of the calibrator
#calno
Serial number of the calibrator
#calcert
Calibrator certificate number
#muldev
Type designation of the multimeter
#mulno
Serial number of the multimeter
#mulcert
Multimeter certificate number
#refdev
Type designation of the reference multimeter
#refno
Serial number of the reference multimeter
#refcert
Reference multimeter certificate number
Procedure Step Parameter
s
Variable Name
Parameter
#fun
(Measuring) function (“V DC”, “RTD” ...)
#mrange
Measuring range (“3.0000 V DC”)
#tpoint
Test point
#outval
Output value and unit of measure (“2.50000 V DC”, “100 °C Pt100” …)
#param
Additional parameter (“1.00 kHz” for V AC, “Pt100” for temperature)
#adig
Permissible deviation in digits
#expval
Expected value
#fexpval
Expected value with measuring function (ACDC)
#minval
Lower limit value
#fminval
Lower limit value with measuring function
#maxval Upper limit value
#fmaxval Upper limit value with measuring function