Operation Manual

GMC-I Messtechnik GmbH 19
GB
1)
Number of display places: 6½ for DC and Ω, 5½ for AC
Resolution is adjustable for the storage and the transmission of
measurement values (see chapter 12.1).
2)
Lowest measurable frequency for sinusoidal measurement signals,
combined measurement of period duration and frequency
3)
At 0 ° ... + 40 °C
4)
As of 10% of the measuring range. See page 20 for influences.
5)
DC component: max. 10% of reading.
6)
ZERO is displayed when the “Zero Balancing” function is active.
7)
Range 100 mV : U
E
= 10 . . . 30 mV
eff
additional error: +0.5% range
1V :U
E
= 0.1 . . .0,3 V
eff
additional error:
+0.3% range
8)
Plus sensor deviation
Measuring
Range
Intrinsic Uncertainty at max. Resolution
at Reference Conditions
Frequency
Range
in Hz
Overload
Capacity
3)
±(...% of rdg. + % of range)
4) 5)
Value Duration
100 mV 0.005 + 0.0006
6)
0.08 + 0.06
7)
45 65
600 Veff
sine
contin-
uous
0.1 + 0.1 10 1 k
5 + 0.5 1 k 5 k
1 V 0.0030 + 0.0004
0.08 + 0.06
7)
45 65
0.1 + 0.1 15 1 k
0.2 + 0.1 10 10 k
5 + 0.5 10 k 50 k
10 V 0.0030 + 0.0004
0.08 + 0.06
0.1 + 0.1
0.2 + 0.1
1 + 0.1
3 + 0.1
45 65
15 1 k
10 10 k
10 k 50 k
50 k 100 k
100 V 0.0030 + 0.0006
600 V 0.0040 + 0.0010
0.08 + 0.06 45 65
0.2 + 0.1 10 1 k
3 + 0.1 1 k 10 k
4) 5)
100 μA
0.02 + 0.002
0.08 + 0.06
0.1 + 0.1
0.2 + 0.1
45 65
10 1 k
1 k 5 k
0.18 A/
600 Veff
contin-
uous
1mA
10 mA
100 mA
±(...% of rdg. + % of range)
100 Ω 0.005 + 0.001
6)
600 Veff
sine
10 min.
1kΩ 0.005 + 0.001
6)
10 kΩ 0.005 + 0.001
100 kΩ 0.005 + 0.001
1MΩ 0.05 + 0.002
10MΩ 0.5 + 0.02
Ω
0.05 + 0.01
1 Hz to 100
kHz
0.05 % of rdg. 600 V
contin-
uous
Pt 100/
Pt 1000
200.0 ... +850.0 °C ±(0.05 % of rdg. + 0.08 K)
8)
600 Veff
sine
10 min.
J 210.0 ... +1200.0 °C
±(0.7 % of rdg. + 0.3 K)
8)
600 Veff
sine
K 270.0 ... + 1372.0 °C