Operation Manual

GMC-I Messtechnik GmbH 13
GB
í Heat or cool the sensor to a known temperature (e.g
0 °C in ice water), and measure the sensor’s resistance
with the Ω
2
function using the 1 kΩ range for Pt100
sensors, and the 10 kΩ range for Pt1000 sensors.
Cable resistance is equal to the difference between the
setpoint (100 Ω from DIN table) and the measured
value.
10.2 Temperature Measurement with Thermocouple
and Reference Junction
Ð Enter the type of thermocouple to be used (J or K) in the
menu mode: rAtE SEnSor ... J K
The reference temperature can either be measured via the
internal reference junction, or can be compensated for
externally, for example through the use of ice water.
Select sensor type and internal reference temperature:
rAtE SEnSor ... IntErn
or
Select sensor type and internal reference temperature in °C:
rAtE SEnSor ... E-tErn
Ð Set the selector switch to “°C
TC
”.
Ð Connect the sensor as shown in the following diagram:
The instrument displays the measured temperature in the unit
of measure which has been entered in the menu mode (param-
eter: “tEMP”).
Note!
The internal reference temperature (from the internal
reference junction) is measured with a temperature
sensor in close proximity to the input jacks. This
temperature is somewhat higher than room temper-
ature due to internal warming. The magnitude of this
deviation has no effect on measuring accuracy.
°
C
TC
+
External Reference Junction: 0°C
°
C
TC
+
Internal Reference Junction
°C
Thermocouple
Equalizing Conductor
Equalizing Conductor
Thermocouple
+