The instrument may only be used for current measurement in
power systems if the electrical circuit is protected with a fuse or a
circuit breaker with a rating of up to 20 A. In order to conform to the
CAT requirements, two fuse links have been fitted for the ranges
mA und A.
Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
Make certain that the measurement cables are in flawless
condition, e.g. no damage to insulation, no interruptions in
cables or plugs etc.
No measurements may be performed with this instrument in
electrical circuits with corona discharge (high voltage).
Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be
present.
Measurements under moist ambient conditions are not per-
mitted.
Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values
can be found in the “Measuring Ranges” table in chapter 4
„Operation“.