User Manual

20 GOSSEN METRAWATT GMBH
METRA HIT 2A: Current Measuring Ranges
Resistance Measuring Ranges
2)
for battery voltage 1.5 V
Capacitance Measuring Ranges
3)
Estimated measurement in the resistance measuring ranges;
determination of measured values via comparative scale, see chapter 4.6.
Current
Voltage Drop approx.
50 µA 0.158 V
1.5 mA 1.16 V 1.21 V
15 mA 1.25 V 1.25 V
150 mA 1.25 V 1.25 V
1.5 A 1.95 V 1.95 V
15 A 0.43 V 0.49 V
Resistance Measuring Limits
Value at Mid-Scale
(R
i
)
Max. Meas. Current
I
max
2)
approx.
x 1
1 Ω ... 1 k
18.00 83 mA
x 10
10 ... 10 k
180.00 8.3 mA
x 100
100 ... 100 k
1.8 0 k 0.83 mA
x 1000
1k ... 1 M
18.00 k 0.083 mA
Capacitance
3)
Measuring Limits
µF x 1000 2000 ... 200000 µF
µF x 100 200 ... 20000 µF
µF x 10 20 ... 2000 µF
µF x 1 2 ... 200 µF