User Manual
GOSSEN METRAWATT GMBH 17
4.7 Measurement of Gain and Attenuation
In communications engineering, gain or attenuation is almost
exclusively given as a logarithm of the ratio between a measured
voltage and a given reference voltage in dB. In recurrent net-
works, it is thereby possible to determine the total gain or atten-
uation in a simple manner by adding or subtracting the individual
values. The reference voltage is 0.775 V (1 mW for 600 Ω);
attenuation at this voltage is 0 dB.
For gain and attenuation measurement, proceed exactly as
described in chapter 4.3.1 for AC voltage measurement; the
measured values, however, are to be read off the dB scale.
The range – 15 ... + 6 dB shown on the scale corresponds to the
AC voltage range 1.5 V. In the case of the higher voltage mea-
suring ranges 5 V , 15 V , 50 V ..., 10 dB, 20 dB, 30 dB ...
are to be added to the value read; see the table in chapter 5
showing the voltage measuring ranges.
If a DC voltage is superposed on the AC voltage to be mea-
sured, this can be cancelled out by using a suitable capacitor,
which is to be series-connected to the measuring input.
The operating voltage of the series-
connected capacitor must be at least
of the same magnitude as the peak
value for the voltage applied. If there is
an additional error of 1% of the measured value, it can be calcu-
lated from the formula opposite.
In mentionned formula, R
i
is the internal resistance of the mea-
suring instrument in the selected measuring range
C
v
1
0,89
f
Hz
-----
R
i
MΩ
-------
⋅⋅
-----------------------------
µF⋅≈