Attention!
!
Only voltage-free objects may be measured. External
voltages falsify the measuring results and might also
damage the instrument!
➭ Read off the value displayed on the Ω scale and multiply by
the factor according to the adjusted measuring range.
If possible, the measuring range should be selected in such a
way as to obtain a reading in the range 5 ... 50 Ω. The measuring
error, in relation to the actual resistance value, is smallest in the
middle of the deflection range. Read off the value displayed on
the Ω scale and multiply by the factor according to the adjusted
measuring range. During resistance measurements of a longer
duration, the full deflection 0 Ω should always be checked if pos-
sible after moving the range switch from one resistance range to
another and, if necessary, it should be reset.
Note!
☞
Contact resistances at the battery connecting terminals
may, particularly in low ohmic resistance ranges, cause
the setting of full deflection 0 Ω to fluctuate. Conse-
quently, good contact should be ensured, for example by
removing and refitting the battery (see chapter 4.2).