User Manual
Edited in Germany • Subject to change without notice • A pdf version is available on the internet.
GMC-I Messtechnik GmbH
Südwestpark 15
90449 Nürnberg•
Germany
Phone +49 911 8602-111
Fax +49 911 8602-777
E-Mail
info@gossenmetrawatt.com
www.gossenmetrawatt.com
18 Guarantee
The guarantee period for all METRA HIT measuring and
calibration instruments is 3 years after date of shipment.
Calibration is guaranteed for a period of 12 months. The
guarantee covers materials and workmanship. Damages
resulting from use for any other than the intended purpose or
operating errors, as well as any and all consequential
damages, are excluded.
19 Product Support
If required please contact:
GMC-I Messtechnik GmbH
Product Support Hotline
Phone +49 911 8602-0
Fax +49 911 8602 709
E-Mail support@gossenmetrawatt.com
20 Recalibration
The respective measuring task and the stress to which your
measuring instrument is subjected affect the ageing of the
components and may result in deviations from the
guaranteed accuracy.
If high measuring accuracy is required and the instrument is
frequently used in field applications, combined with transport
stress and great temperature fluctuations, we recommend a
relatively short calibration interval of 1 year. If your measuring
instrument is mainly used in the laboratory and indoors
without being exposed to any major climatic or mechanical
stress, a calibration interval of 2-3 years is usually sufficient.
During recalibration* in an accredited calibration laboratory
(DIN EN ISO/IEC 17025) the deviations of your instrument in
relation to traceable standards are measured and
documented. The deviations determined in the process are
used for correction of the readings during subsequent
application.
We are pleased to perform DAkkS or factory calibrations for
you in our calibration laboratory. Please visit our website at
www.gossenmetrawatt.com ( Company DAkkS
Calibration Center or FAQs Calibration questions and
answers).
By having your measuring instrument calibrated regularly, you
fulfill the requirements of a quality management system per
DIN EN ISO 9001.
* Verification of specifications or adjustment services are not part of the
calibration. For products from our factory, however, any necessary adjustment
is frequently performed and the observance of the relevant specification is
confirmed.