7.2 Frequency Measurement [Hz]
Ð Select the V~/Hz measuring function with the rotary
selector switch.
Ð Apply the measured quantity as described under voltage
measurement.
Ð Briefly press the FUNC key in order to measure frequency.
The frequency values appears at the main display. The
frequency measuring range can be selected
subsequently by pressing MAN|AUTO. The previously
selected voltage measuring range remains active.
Ð The instrument can be switched from frequency
measurement back to alternating voltage measurement
by pressing the FUNC key once again. This selection is
acknowledged with an acoustic signal.
Note!
Frequency measurement is only possible if the
measuring signal passes through zero (AC coupling).
8 Resistance and Diode Measurements
8.1 Resistance Measurement (2-wire connection) []
Ð Make sure that the device under test is voltage-free (see
chapter 7.1). Interference voltages distort measurement
results and damage the instrument!
Ð Set the rotary selector switch to “”.
Ð Connect the device under test as shown.
Note!
High Impedance Measurements in the 3 M/ 30 M
Range
In the event of extremely high resistances the
capacitive influence of the person performing the
measurement and/or the measuring cable may
distort the measured value.
Use therefore short or shielded measurement cables.
8.1.1 Zero Balancing in the 300 and 3 k Measuring Ranges
Cable and contact resistances can be eliminated for the
measurement of small resistance values in the 300 and
3k ranges by means of zero balancing:
Ð Plug the measuring cables into the instrument and
connect the free ends to each other.
Ð Press the FUNC key.
The instrument acknowledges zero balancing with an
acoustic signal, and „000.00 “ or „0.0000 k“ and the “ZERO”
symbol appear at the display. The resistance value measured
at the moment the key is pressed serves as a reference value
(max. 20 , or 200 , or 2000 digits). It is automatically
subtracted from all subsequent measured values. If the
measuring range is changed (MAN|AUTO key), the ZERO
function remains active (at the display and in memory).
Ð Zero balancing can be deleted:
– By once again pressing and holding the FUNC key, which
is acknowledged with two acoustic signals,
– By switching the instrument off.