User manual

12 GMC-I Messtechnik GmbH
8.2 Continuity Test during Resistance Measurement
If the “acoustic signal ” function is activated and the
0 ... 310 measuring range is selected, a continuous
acoustic signal is generated by the instrument within a range
of 0 to approx. 10 .
Overflow “0l is displayed where Rd > 310 .
Activating and Deactivating Continuity Testing (acoustic signal)
Set the rotary selector switch to „//.
Repeatedly press the FUNC key until and “short” appear
at the display, provided the connector jacks are open
(Display 0.L).
Connect the measurement cables to the device under
test.
The instrument is switched to resistance measurement
by pressing the FUNC key once again.
8.3 Diode Measurements
Make sure that the device under test is voltage-free (see
chapter 7.1). Interference voltages distort measurement
results!
Set the rotary selector switch to „//.
Briefly press the FUNC key for diode measurement. Unit of
measure “V” and “diode” are then displayed if the
connector jacks are open (Display 0.L).
Connect the device under test as shown.
Conducting Direction and Short-Circuit
The instrument displays conducting-state voltage in volts. As
long as voltage drop does not exceed the maximum display
value of 3 V, several series connected components or
reference diodes can be tested.
Reverse Direction or Interruption
Overload (.0l) is displayed at the instrument for diode
measurements where Ud > 3.1 V. Measuring current is
always a constant current of approximately 1 mA.
Note!
Resistors and semiconductors which are connected
in parallel to the diode distort measurement results!
m
S+S–
VC
Conducting Direction
Reverse Direction