8 Resistance and Diode Measurements
8.1 Resistance Measurement (2-wire connection) []
Make sure that the device under test is voltage-free (see
chapter 7.1). Interference voltages distort measurement
results and damage the instrument!
Set the rotary selector switch to “”.
Connect the device under test as shown.
Note!
High Impedance Measurements in the 3 M/ 30 M
Range
In the event of extremely high resistances the
capacitive influence of the person performing the
measurement and/or the measuring cable may
distort the measured value.
Use therefore short or shielded measurement cables.
8.1.1 Zero Balancing in the 300 and 3 k Measuring Ranges
Cable and contact resistances can be eliminated for the
measurement of small resistance values in the 300 and
3k ranges by means of zero balancing:
Plug the measuring cables into the instrument and
connect the free ends to each other.
Press the FUNC key.
The instrument acknowledges zero balancing with an
acoustic signal, and „000.00 “ or „0.0000 k“ and the “ZERO
symbol appear at the display. The resistance value measured
at the moment the key is pressed serves as a reference value
(max. 20 , or 200 , or 2000 digits). It is automatically
subtracted from all subsequent measured values. If the
measuring range is changed (MAN|AUTO key), the ZERO
function remains active (at the display and in memory).
Zero balancing can be deleted:
By once again pressing and holding the FUNC key, which
is acknowledged with two acoustic signals,
By switching the instrument off.
0.L
k
FUNC
0.5435
diode
V
FUNC
000. .0
short
FUNC
m
S+S–
VC
R
x
Voltage Drop
: 10 m 30 M
Overall measuring range:
+