User Manual

GMC-I Messtechnik GmbH 13
12 Characteristic Values
1)
At 0 C...+40C
2)
with zero balancing, + 35 digits without zero balancing
3)
without sensor
4)
12 A 5 minutes, 16 A 30 seconds
5)
1 ... 35 d from the zero point due to TRMS converter when probe tips are short-
circuited
6)
power limiting: frequency x voltage max. 6 x 106 V x Hz for U > 100 V
Key
rdg. = reading (measured value)
d = digit
Meas.
Function
Measuring Range
Reso-
lution
Input Impedance
Intrinsic Error at Max. Resolution
under Reference Conditions
Overload Capacity
1)
Meas.
Function
(... % rdg. + ... d) (... % rdg. + ... d)
6000
5)
Value Time
V
600 mV 100 V>10G //<40pF 40M // < 40 pF 0.5 + 5
1 + 5
600 V
DC
AC
eff
Sinus
Cont.
V
6V 1mV11M // < 40 pF 8 M // < 40 pF 0.5 + 5
60 V 10 mV 10 M // < 40 pF 8 M // < 40 pF 0.5 + 5
600 V 100 mV 10 M // < 40 pF 8 M // < 40 pF 0.5 + 5
Voltage drop at approx. range limit
5)
A
60 mA 10 A 100 mV 100 mV 1.0 + 5 (> 10 D) 1.5 + 5 (> 10 D)
1,0 A Cont.
A
600 mA 100 A 700 mV 700 mV 1.0 + 5 1.5 + 5 (> 10 D)
6 A 1 mA 200 mV 200 mV 1.0 + 5 (> 10 D) 1.5 + 5 (> 10 D)
10 A
4)
Cont.
10 A 10 mA 300 mV 300 mV 1.0 + 5 1.5 + 5 (> 10 D)
Open-circuit voltage
Meas. current at
range limit
(... % rdg. + ... d)
600 100 m max. 1 V max. 250 A1 + 5
2)
600 V
DC
AC
eff
Sinus
max. 10 s
6k 1 max. 1 V max. 100 A 0.7 + 3
60 k 10 max. 1 V max. 12 A 0.7 + 3
600 k 100 max. 1 V max. 1.2 A 0.7 + 3
6M 1k max. 1 V max. 120 nA 0.7 + 3
40 M 10 k max. 1 V max. 50 nA 2.0 + 3
2V 1mV max. 3V 1 + 5
(... % rdg. + ... d)
600 0,1 < 1 V 1 + 5
600 V
DC
AC
max. 10 s
(... % rdg. + ... d)
C
TYP K
50.0 ...
+400.0 C
0.1 C 1.0 + 5
3)
600 V
DC/AC
eff
Sinus
max. 10 s
C
(... % rdg. + ... d)
Hz
100 Hz
0,1 Hz 0.1 + 2
600 V
6)
Hz
1000 Hz
1 Hz 0.1 + 2