Operation Manual

2 GOSSEN-METRAWATT
Specifications METRA
Hit
12S ... 16S, 14A
1) TRMS measurement
2) Direct display with clip-on transformer 1000:1
4) At 0 °C...+40°C
5) With zero setting; w/o zero setting + 35 digits
6) With zero setting; w/o zero setting + 50 digits
7) METRAHit 13S (w/o 16 A fuse!): 16 A cont., 20 A for 5 min;
METRAHit 14A, 14S ... 16S: 12 A for 5 min, 16 A 30 s
8) Range 3 V : U
E
= 1.5 V
rms
. . . 100 V
rms
30 V : U
E
= 15 V
rms
. . . 300 V
rms
300 V : U
E
= 150 V
rms
. . . 1000 V
rms
9) On the range 3V , rectangular signal positive at one end 5 ... 15 V, f = const.,
not 163.84 Hz or integer multiple.
10) Without sensor
Meas.
function
Measuring range
Resolution Input impedance
Inherent deviation of the digital display
±(...% of meas. val. +... digits) for reference conditions
Overload capacity
4)
Measuring
function
Overload
value
Overload
duration
METRA
Hit
12S 13S 14A 14S 15S 16S 12S 13S/14A 14S 15S 16S
V
30.00 mV
●●●●
10 µV>10G//<40pF 0.5 + 3
5)
0.5+3
5)
1200 V
DC
AC
effective
sinusoidal
cont.
V
300.0 mV
●●●●
100 µV>10G//<40pF 0.5 + 3 0.5+3
3.000 V
●●●●
1mV 11M//<40pF 0.25 + 1 0.1+1
30.00 V
●●●●
10 mV 10 M//<40pF 0.25 + 1 0.1+1
300.0 V
●●●●
100 mV 10 M//<40pF 0.25 + 1 0.1+1
1000 V
●●●●
1 V 10 M //<40pF 0.35 + 1 0.1+1
V
3.000 V
●●●●
1)
1mV 11M//<40pF
0.75 + 2 (10 ... 300 D)
0.75 + 1 (> 300 D)
0.75 + 3
(> 10 D)
V
30.00 V
●●●●
1)
10 mV 10 M//<40pF
300.0 V
●●●●
1)
100 mV 10 M//<40pF
1000 V
●●●●
1)
1 V 10 M//<40pF
V
3.000 V
1)
1mV 11M//<40pF
0,75 + 3
(> 10 D)
V
30.00 V
1)
10 mV 10 M//<40pF
300.0 V
1)
100 mV 10 M//<40pF
1000 V
1)
1 V 10 M//<40pF
Voltage drop, approx.
12S 13S /14A
14S /
15S/16S
A
300.0 µA
●●
100 nA ——15 mV 1.0 + 5 (> 10 D) 0.5 + 5 (> 10 D)
0.36 A cont.
A
3.000 mA
●●●●
1 µA 15 mV 15 mV 150 mV 1.0 + 5 (> 10 D) 1,0+2 0.5+2
30.00 mA
●●●●
10 µA 150 mV 150 mV 650 mV 0.25 + 2 1.0 + 5 (> 10 D) 0.5 + 5 (> 10 D)
300.0 mA
●●●●
100 µA 1V 1V 1V 1.0 + 2 0,5+2
3.000 A
●●●●
1mA 100 mV 100 mV 1.0 + 5 (> 10 D) 1.0 + 5 (> 10 D)
7) 7)
10.00 A
16A ●●●●
10 mA 300/270mV 270 mV 1.0+2 1.0+2
A
3.000 mA
1 µA ——150 mV 1.5 + 2 (> 10 D)
0.36 A cont.
A
30.00 mA
10 µA 150 mV 150 mV 1.5 + 2 (> 10 D)
300.0 mA
●●
100 µA 1V 1V 1V 1.5 + 2 (> 10 D)
7) 7)
10.00 A
16A ●●
10 mA 300/270mV 270 mV 1.5 + 2 (> 10 D)
A
30.00 A
2)
10 mA 150 mV
1.5+2
(> 10 D)
0.36 A cont.
A
300.0 A
2)
100 mA 1V
A
3.000 mA
1)
1 µA ——150 mV 1.5 + 4 (> 10 D)
A300.0 mA
1)
100 µA ——1V 1.5 + 4 (> 10 D)
12 A 5min
10.00 A
1)
10 mA ——270 mV 1.75 + 4 (>10 D)
No-load voltage
30.00
●●●●
10 m max. 3.2 V 0.5 + 3
5)
0.4+3
5)
500 V
DC
AC
effective
sinusoidal
10 min
300.0
●●●●
100 m max. 3.2 V 0.5 + 3 0.4+3
3.000 k
●●●●
1 max. 1.25 V 0.4 + 1 0.2+1
30.00 k
●●●●
10 max. 1.25 V 0.4 + 1 0.2+1
300.0 k
●●●●
100 max. 1.25 V 0.4 + 1 0.2+1
3.000M
●●●●
1k max. 1.25 V 0.6 + 1 0.4+1
30.00M
●●●●
10 k max. 1.25 V 2.0 + 1 2.0+1
2.000 V
●●●●
1mV max. 3.2 V 0.25 + 1 0.1+1
Discharge
resistance
U
0 max
F
30.00 nF
●●
10 pF 250 k 2.5 V 1.0+3
6)
500 V
DC / AC
effective
sinusoidal
10 min
F
300.0 nF
●●
100 pF 250 k 2.5 V 1.0+3
3.000 µF
●●
1nF 25 k 2.5 V 1.0+3
30.00 µF
●●
10 nF 25 k 2.5 V 3.0+3
Sensor f
min
V f
min
V
Hz
300.0 Hz
●●
0.1 Hz 1 Hz 45 Hz
0.5+1
8)
3 kHz:
1200 V
30 kHz:
300 V
100 kHz:
30 V
cont.
Hz
3.000 kHz
●●
1Hz 1Hz 45Hz
30.00 kHz
●●
10 Hz 10 Hz 45 Hz
100.0 kHz
●●
100 Hz 100 Hz 100 Hz
%
2.0 ... 98.0 %
●●
0.1 % 1 Hz
1 Hz...1 kHz: ±5 D
9)
1 kHz...10 kHz: ±5 D/kHz
9)
%
°C
– 200.0 ...
+ 200.0 °C
●●●●
0.1 °C
Pt 100
2 Kelvin + 5 D
10)
500 V
DC
AC
effective
sinusoidal
10 min
°C
+ 200.0 ...
+ 850.0 °C
●●●●
0.1 °C 1.0+5
10)
– 100.0 ...
+ 200.0 °C
●●●●
0.1 °C
Pt 1000
—— 2 Kelvin + 2 D
10)
+ 200.0 ...
+ 850.0 °C
●●●●
0.1 °C 1.0+2
10)
Analog-digital multimeters
METRA 12S ... 18S, 14A