Datasheet
METRACLIP
61
Digital Current Clamp Meter
3-349-174-03
5/4.17
GMC-I Messtechnik GmbH
Features
DATA HOLD Function
For measurements under unfavourable conditions, the momen-
tary measured value can be „frozen“ at the display.
Signalling in the Event of Over-Ranging
There is an optical signal for over-ranging.
Battery Saving Circuit
The device turns itself off automatically 10 minutes after it has
been switched on.
Applicable Regulations and Standards
Characteristic Values
Alternating Current mA AC/A AC (Manually Selectable)
Frequency range 50 … 60 Hz
Measuring principle arithmetic mean
Reference Conditions
Ambient temperature +23 °C ±5 °C
Relative humidity max. 80%
IEC 61010-1/EN 61010-1/
VDE 0411-1
Safety requirements for electrical equipment for
measurement, control and laboratory use
EN 60529
VDE 0470 Part 1
Test instruments and test procedures
Protection provided by enclosures (IP code)
IEC 61326/EN 61326 Electromagnetic compatibility (EMC)
Measuring
Range
Range of
Application
Resolution Accuracy
0 … 30 mA
0 … 300 mA
0 … 300 mA
0.01 mA
0.1 mA
±1.2% rdg. ±5 digits
0 … 30.00 A
0 … 300.0 A
0 … 200 A
0.01 A
0.1 A
±1.2% rdg. ± 5 digits
200 … 250 A ± 3% rdg. ± 5 digits
250 … 300 A ± 5% rdg. ± 5 digits
• Measurement of leakage current and load current
• Display digital and via bargraph
• Two measuring ranges
convertible between 30 mA/300 mA and 30 A/300 A
• Insensitive to external magnetic fields
• Large clamp opening with a diameter of 40 mm
• Compact, handy und lightweight