Datasheet

Digital Luxmeter
MAVOLUX 5032 C USB / MAVOLUX 5032 B USB
Calibration Certificate – optional at additional price
Calibration reference: Scientific Standard Lamp, type Wi 41 G of the
PTB, (Physikalisch Technische Bundesanstalt Braunschweig –
National Standards Institute). This Calibration Certificate is
necessary for all applications, where measurements, controls and
inspections are required by the relevant laws and regulations.
Technical Data MAVOLUX 5032 C USB
Meas.
Quantity
Measuring
Range
in Lux (lx)
Meas. Range in
footcandle (fc)
Reso-
lution
in Lux
Reso-
lution
in fc
I 0.1 ... 199.9 0.01 ... 19.99 0.1 0.01
II 1 ... 1 999 0.1 ... 199.9 1 0.1
III 10 ... 19 900 1 ... 1 999 10 1
Illumi-
nance
IV 100 ...199 000 10 ... 19 990 100 10
in candela/m² in footlambert in
cd/m²
in fL
I 1 ... 1 999 0.1 ... 199.9 1 0.1
II 10 ... 19 990 1 ... 1 999 10 1
III 100 ... 199 900 10 ... 19 990 100 10
Lumi-
nance
(with
attach.)
IV 1000 .1 999 000 100 ...199 900 1000 100
Technical Data MAVOLUX 5032 B USB
Meas.
Quantity
Measuring
Range
in Lux (lx)
Meas. Range in
footcandle (fc)
Reso-
lution
in Lux
Reso-
lution
in fc
I 0.01 ... 19.99 0.001 ... 1.999 0.01 0.001
II 0.1 ... 199.9 0.01 ... 19.99 0.1 0.01
III 1 ... 1 999 0.1 ... 199.9 1 0.1
IV 10 ... 19 900 1 ... 1 999 10 1
Illumi-
nance
V 100 ... 199 000 10 ... 19 990 100 10
in candela/m² in footlambert in
cd/m²
in fL
I 0.1 ... 199.9 0.01 ... 19.99 0.1 0.01
II 1 ... 1 999 0.1 ... 199.9 1 0.1
III 10 ... 19 999 1 ... 1 999 10 1
Battery Saver Circuit
The instrument is automatically switched off, if the
measurement value remains unchanged for approximately
4 minutes, and if none of the operating elements has been
activated during this time.
Continuous operation
For using the instrument in the continuous operation mode,
keep the key HOLD pushed down and then press the key
ON/OFF. Pushing the key ON/OFF again will end the
operation
Electromagnetic Compatibility EMC
The MAVOLUX meet the Specifications 89/336/EWG dt.
01.01.1996
Mechanical Design
Dimensions Basic instrument 65 x 120 x 19 mm
Sensor 31 x 105 x 30 mm
Cable between Instrument
and Sensor MAVOLUX 5032 C-USB: coiled cord,
permanently attached
MAVOLUX 5032 B-USB: plug-in coiled
cord.
Cable Length 1,5 m; available also in 3 m, 5 m or
10 m – on special order
Weight approx. 200 g without battery
Standard Equipment
1 Battery 1,5 V
1 Instruction manual
1 USB – Cable
1 CD-ROM with software for processing the values measured
and regulating the measuring instrument
1 ever-ready case for luxmeter
Optional Accessories
1 Luminance attachment cd/m² incl. case
1 Luminance attachment fL incl. case
1 Calculator, calculating disc for exposure time/aperture
combinations and additional values for photography
IV 100 ... 199 900 10 ... 19 990 100 10
Lumi-
nance
(with
attach.)
V 1000 ..1999 000 100 ...199 900 1000 100
Ordering Numbers
_____________________________________________________________________________________________________________________
GOSSEN Foto- u. Lichtmesstechnik GmbH Page 3
Description Order No. Price
MAVOLUX 5032 C USB - according to DIN 5032-7 Class C
incl. ever-ready case, USB - cable, Software
M502G
MAVOLUX 5032 B USB - according to DIN 5032-7 Class B
incl. ever-ready case, USB - cable, Software
M503G
Luminance attachment, cd/m² with case 5908V0120
Luminance attachment fL with case Z481B
Calculator in case 5999V0380
Calibration certificate for MAVOLUX 5032 C/B-USB H997B
Optional: Special lengths of connecting cable: 3 or 5 or 10 meters please call