Datasheet

11
Specifications
Constant lighting conditions A green LED indicates adherence to the
admissible indoor light range for diagnostics, and a red LED indicates
violation of this range.
Reduced testing effort The Surveillance of ambient light extends the
standards-specified interval for constancy testing of image display
devices for veiling luminance and maximum contrast to 6 months.
In the case of repeat measurements, there is no need to repeat the
veiling luminance measurement.
Selectable indoor light range Two variants are available: one for general
applications MAVOMAX 60 (20 to 60 lx), and one for use in the field of
mammography MAVOMAX RK1 (10 to 50 lx) (corresponds to RK1 per draft
standard DIN 6868-157).
Flexible power supply Electrical power for the continuous operating mode
is supplied to the meter by connecting it to a free USB port or the included
mains power pack.