The surveillance of the ambient light extends the standards-
specified interval for constancy testing of image display devices for veiling
luminance and maximum contrast to 6 months. In the case of repeat
measurements within the framework of acceptance or constancy testing,
veiling luminance need not be measured again, and the 60-minute waiting
time until the image display device has reached a stable state is eliminated.
Further applications would also be the surveillance of the conditions for
dimmable light as well as for external workstations for emergency diagnostics.
MAVOMAX
completes the application of every diagnostic monitor or light box for medical applications according to norm DIN 6856-1. It guarantees consistent lighting and
performs light surveillance for monitors in accordance with norm DIN EN IEC 61223-2-5 (QS-RL dated 20 Nov. 2003) and DIN V 6868-57.