Device Under Test
Connect DUT to the test socket.
Connect DUT to mains outlet.
– Switch the DUT on.
Protective
conductor
resistance
Probe socket
PE/ I
C
Protective
conductor
resistance
Permanent,
probe socket
PE/ I
C
Insulation
resistance
No probe
required
Insulation
resistance
Probe
socket R
ISO
Differential
current
No probe
required
Contact current
Probe socket
PE/ I
C
Safety class I
Earth contact plug
All exposed
parts are con-
nected with PE
✘✘
Test with the
mains plug
poled in both
directions.
With exposed
parts which
are not con-
nected with PE
Test with the
mains plug
poled in both
directions.
Test with the
mains plug
poled in both
directions.
CEE plug *
All exposed
parts are con-
nected with PE
✘✘✘
With exposed
parts which
are not con-
nected with PE
?
✘✘
Permanently
connected DUT
✘✘
Safety class II
With European
earth contact plug
Test with the
mains plug
poled in both
directions.
Use of mains outlet number depending on the mains terminal
(all test sockets as well as mains outlets No. 8 and 19 can always be used
regardless of which mains adapters are used)
Mains outlet number
20 21 22
Mains connection with mains plug 3P+N+PE 32 A
✘✘✘
Connection with earthing contact plug adapter to CEE coupling 3P+N+PE 32 A-6h
max. 16 A
Connection with adapter for plug 1P+N+PE 16 A to CEE coupling 3P+N+PE 32 A-6h
max. 16 A
Connection with adapter for plug 3P+N+PE 16 A to CEE coupling 3P+N+PE 32 A-6h
max. 16 A
max. 16 A
Connection with adapter for plug 1P+N+PE 32 A to CEE coupling 3P+N+PE 32 A-6h