User manual
GMC-I Messtechnik GmbH 7
* MINITEST 3P Master only
Device Under Test
Connect DUT to the test socket.
Connect DUT to mains outlet.
– Switch the DUT on.
Protective
conductor
resistance
Probe socket
PE/ I
C
Protective
conductor
resistance
Permanent,
probe socket
PE/ I
C
Insulation
resistance
No probe
required
Insulation
resistance
Probe
socket R
ISO
Differential
current
No probe
required
Contact current
Probe socket
PE/ I
C
Safety class I
Earth contact plug
All exposed
parts are con-
nected with PE
✘✘
✘
Test with the
mains plug
poled in both
directions.
With exposed
parts which
are not con-
nected with PE
✘
✘
Test with the
mains plug
poled in both
directions.
✘
Test with the
mains plug
poled in both
directions.
CEE plug *
All exposed
parts are con-
nected with PE
✘✘✘
With exposed
parts which
are not con-
nected with PE
✘
?
✘✘
Permanently
connected DUT
✘✘
Safety class II
With European
earth contact plug
✘
✘
Test with the
mains plug
poled in both
directions.
Use of mains outlet number depending on the mains terminal
(all test sockets as well as mains outlets No. 8 and 19 can always be used
regardless of which mains adapters are used)
Mains outlet number
20 21 22
Mains connection with mains plug 3P+N+PE 32 A
✘✘✘
Connection with earthing contact plug adapter to CEE coupling 3P+N+PE 32 A-6h
–
✘
max. 16 A
–
Connection with adapter for plug 1P+N+PE 16 A to CEE coupling 3P+N+PE 32 A-6h
–
✘
max. 16 A
–
Connection with adapter for plug 3P+N+PE 16 A to CEE coupling 3P+N+PE 32 A-6h
✘
✘
max. 16 A
✘
max. 16 A
Connection with adapter for plug 1P+N+PE 32 A to CEE coupling 3P+N+PE 32 A-6h
– ✘ –