Datasheet
3-349-357-03
7/1.11
GMC-I Messtechnik GmbH
MINITESTMASTERPROBASE
Testers per DIN VDE 0701-0702
MINITESTMASTERPRO
MINITESTBASE
MINITEST 3PMASTER
Features
* as from series issued in March 2007
Convenient Connection
The test instrument is intended for testing and measuring repaired
or modified devices. The device under test is connected to the
test instrument’s test socket to this end.
When testing protective conductor current and contact current
(absence of voltage at exposed, conductive parts), the device
under test is connected to the mains outlet on the test instrument.
Display Functions
Limit value violations are indicated optically by means of nine
variously colored LEDs.
MINI
TEST
MASTERPRO: All measured values are also clearly read
out at a large, two-line digital display.
Rugged Mechanical Design
The handy instrument is furnished with a compact plastic housing
with permanently connected mains cable. The respective mea-
sured quantity is selected by means of a rotary switch.
PC Analysis Software (not for MINITESTB
ASE)
The measurement data can be transferred to a PC for onward
processing with one of our software packages.
Features of MINITEST... series
MINITEST 3PMASTER
MINITESTMASTER
MINITESTPRO
MINITESTBASE
Connection types
Power supply via permanently connected mains cable 8888
Tests on monophase DUTs 8888
Tests on 3-phase DUTs via additional test sockets
CEE 16A / CEE 32A
8 –––
Fusing devices
Fuse for probe connection 8888
RCCB in mains plug – 88–
Miniature circuit breaker 8 –––
Protocol functions
Illuminated two-line LCD * 888 –
Memory for 2,000 tests (10 measured values per test) 88 ––
Key for transmission of measured values 888 –
Key for storing measured values 88 ––
Data interface (USB port) 888 –
Barcode scanner connection (9 pin, subminiature plug)
for reading ident. numbers in text form with a maximum of 24
characters as description of DUT
88 ––
Applications
Testing the electrical safety of electrical equipment
per DIN VDE 0701-0702: 2008
by measuring:
• protective conductor resistance
• insulation resistance
• protective conductor current –
differential current method
• contact current –
direct measurement method
• absence of voltage by means of