User manual
58 GMC-I Messtechnik GmbH
Classification Parameter – VDE 0701-0702-PRCD
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
3
New Classification Parameters „PRCD Type“
(as from Firmware V1.7.0)
(are only displayed if parameter „Standard VDE 0701-0702-
PRCD“ has been selected):
– PRCD (Standard):
For the testing of simple circuit breaker safety adapters in
which the protective conductor is permanently connected.
– PRCD (SPE):
(SPE = Switched Protective Earth) For the testing of PRCDs,
in which the protective conductor is only connected in
switched-on condition.
– PRCD-S (SPE):
For the testing of circuit breaker safety adapters of type
PRCD-S.
– PRCD-K (SPE):
For the testing of circuit breaker safety adapters of type
PRCD-K.
Note
The standard or standard variant associated with the
respective selector switch position conforms to the
default setting.
Ax means that the standard variant VDE 0701-0702-
PRCD can be selected in each of the preset switch posi-
tions.
Note
For more information on the testing of single-phase and
3-phase type S and K PRCDs by simulating faults, please
visit our company website under item PROFITEST PRCD test
adapter.
Note
The testing of PRCDs (test sequences and time to trip) is
only possible for DUTs with a nominal voltage of 230 V.
Classification Parameter – IEC 62353
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT, see table above
VDE 0701-0702-ExtC, see previous table
VDE 0701-0702-PRCD
2
IEC 62353, see following table
IEC 60974-4, see table below
Protection class
1 2
Class I, Class I+II
Connection
1 2
test socket
2/2
Meas. Mode
1
PRCD
3
PRCD Typ
3
PRCD (Standard)
PRCD (SPE)
PRCD-S (SPE)
PRCD-K (SPE)
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT, see table above
VDE 0701-0702-ExtC, see table above
VDE 0701-0702-PRCD, see previous table
IEC 62353
IEC 60974-4, see following table
Protection class
1
Class I, Class II or Class I+II
Connection type
1
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: AT3-Adapter
Perman. Con.: P1+P2 (only with feature H01)
2/2
Measurement type (MA)
1
passive
active
Applied parts Applied parts: none, B, BF, CF or combinations
Type B (body): Devices of this type are suitable for both
internal and external patient applications, except for use in
direct proximity to the heart.
The following protection classes are permissible:
I, II, III or devices with internal electrical power supply.
Type BF (body float): same as type B, but with type F in-
sulated applied parts.
Type CF (cardiac float): Devices of this type are suitable
for use directly at the heart. The applied part may not be
grounded.
The following protection classes are permissible:
I, II or devices with internal electrical power supply.
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protec-
tion class must be entered manually.