Change Classification Parameters Conveniently
(Option Feature E01 (Touch Screen))
Per touch click (brief tapping) on the respective window of the
classification parameters the corresponing selection menu
opens.
You automatically go back to the start menu by selecting the
desired parameter.
Classification Parameter – VDE 0701-0702
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
Classification Parameter – VDE 0701-0702-IT
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
Classification Parameter – VDE 0701-0702-ExtC
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section (with EL1 only length). Data remain in
memory until a new entry is made.
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence
VDE 0701-0702
VDE 0701-0702-IT, see following table
VDE 0701-0702-ExtC, see table below
VDE 0701-0702-PRCD, see table below
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I, Class II, Class I+II, Class I+III, Class II+III, Class I+II+III
Connection
1 2
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: VL2E-Adapter
Adapter: AT3-Adapter (feature I01)
Perman. Con.: P1+P2 (only with feature H01)
2/2
Meas. Mode
1)
passive
active
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT
VDE 0701-0702-ExtC, see table below
VDE 0701-0702-PRCD, see table below
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I, Class II, Class I+II, Class I+III, Class II+III, Class I+II+III
Connection
1 2
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: AT3-Adapter (feature I01)
2/2
Meas. Mode
1
active
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT, see previous table
VDE 0701-0702-ExtC
VDE 0701-0702-PRCD, see following table
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I
Connection
1 2
test socket
Adapter: AT3-I IIE-Adapter
Adapter: EL1-Adapter
Adapter: VL2E-Adapter
2/2
Meas. Mode
1
ExtCord
2
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.