Wiring Diagram
Alternative Measuring Method (equivalent leakage current)
Alternative measurement type
– DUT mains plug (protection classes I) to test socket
Schematic Diagram
After activating test voltage, leakage current is measured via the
DUT’s mains cable between short-circuited mains conductors L
and N and the protective conductor terminal at the DUT.
Wiring Diagram
Connection of 3-phase DUTs (only with SECUTEST PRO or feature
I01 with optional test adapter AT3-IIIE)
AT3-Adapter measurement type
– DUT mains plug to AT3-IIIE test adapter
– AT3-IIIE probe to COM-V terminals
– AT3-IIIE test plug to test socket
Schematic Diagram
Measurement of the DUT with 3-phase mains connection via
AT3-IIIE adapter
Wiring Diagram (AT3-IIIE probe to COM-V )
Measurement of protective conductor current via current clamp
sensor with voltage output for permanently installed DUTs (only
with SECUTEST PRO or feature I01 with optional WZ12C current
clamp sensor)
Clamp measurement type
Schematic Diagram
Measurement of protective conductor current by closing the cur-
rent clamp sensor around mains PE in the power cable for per-
manently installed protection class I devices under test.
Wiring Diagram (Current Clamp Sensor to COM-V)
Set Measuring Range at Current Clamp Sensor and Parameter at
the SECUTEST PRO
* only with WZ12C
** Default value
SECUTEST PRO
Current Clamp Sensor SECUTEST PRO
Transformer Ratio
Parameter
Transformer Ratio
(Switch*)
Measuring
Range
Display Range
with Clamp
1 mV : 1 mA
WZ12C
1 mV : 1 mA 1 mA... 15 A 0 mA ... 300 A
100 mV : 1 mA **De-
fault value
SECUTEST CLIP
100 mV : 1 mA 0.1...25 mA 0.00 mA ... 3.00 A