User manual

20 GMC-I Messtechnik GmbH
7 Operation with Barcode Reader
The barcode reader Z720A (as accessory) allows for all information avail-
able in barcode form to be quickly, easily and safely entered into the test
reports. This type of data entry permits time-saving and low-cost acqui-
sition of large amounts of information, e.g. for series measurements of
devices provided with barcodes.
7.1 Connecting the Barcode Reader
Connect the reader to the RS232 interface of the SI module.
An SI LCD window must not be active!
The barcode reader confirms the correct connection with a double
acoustic signal.
7.2 Configuring the Barcode Reader
The barcode reader Z720A is configured for the following barcodes:
CODE 39 / CODE 128 / EAN13 (12 digits) *
The barcode reader is immediately ready for operation when connected
to a SECUTEST... or SECUTEST SI+ test instrument.
For operation with PROFITEST 204, the barcode reader has to be con-
figured with the code indicated in the operating instructions of the bar-
code reader adapter. For these test instruments, only Code 128 is possi-
ble.
If you wish to use other codes for the SECUTEST..., please consult our
Product Support Hotline, see chapter 13, page 26
* Z720A has the appropriate scanning width to be programmed to the EAN 128
barcode.