Electrical Safety
Ambient Conditions (EN 61010-031)
Temperature 20 °C ... + 50 °C
Relative humidity Max. 80%
Pollution degree
2
Using the KS17-4
Attention!
!
Observe the instrument’s maximum values for electrical safety.
Measurements per DIN EN 61010-031 may only be per-
formed in environments in accordance with measuring
categories III and IV with the safety cap attached to the
test probe at the end of the measurement cable.
In order to establish contact inside 4 mm jacks, the safety caps
have to be removed by prying open the snap fastener with a
pointed object (e.g. the other test probe).
1
Bottom Right
Detach the strap from the instrument:
Turn out the slotted screw (M3) at the
bottom.
Bottom Left
Strap
Clasp
Clasp
Front View
2
Slide the strap through the test probe
holder.
Eyelet for Attachment to Tester
Side View
(METRISO INTRO, BASE, TECH, PRO, XTRA)
3
Front
Feed the strap through from the front
of the test instrument and secure it
with the slotted screw (M3).
Back
Maximum rated voltage 600 V
1000 V
1000 V
Measuring category
CAT IV CAT III CAT II
Max. rated current: 1 A 1 A 16 A
With safety cap attached
••
Without safety cap ——