11 Characteristic Values
METRISO XTRA
1)
the indicated accuracy is only achieved with the optional accessory
„shielded high-resistance measuring cable KS-C (article number Z541F)“
2)
does not conform to DIN EN 61557-2
3)
Display range up to 1.2 kV
4)
up to 5 Ω
Breakdown Voltage (Uramp)
Polarization Index (PI), Absorption Ratio (DAR)
PI and DAR are calculated values. Insulation measurement specifications apply.
Reference Conditions
Reference
temperature + 23 °C ±3 K
Relative humidity 40 to 75%
Measured quantity
frequency 45 to 65 Hz
Measured quantity
waveshape Sine, deviation between TRMS and recti-
fied value < 1%
Battery voltage 9.5 V ±0.1 V
Test resistance 10 MΩ±1%
Electrical Safety
Protection class II per IEC/EN 61010-1/VDE 0411-1
Pollution degree 2
Measuring category
CAT II 1000 V / CAT III 600 V / CAT IV 300 V
Fuses
Fuse link FF315mA/1000V, effective in all resis-
tance measuring ranges, 1 additional
replacement fuse in the battery compart-
ment
Electronic fuse Protects low-resistance and resistance
measurements R
LO
and R
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2013, class B
Interference immunity EN 61326-1:2013
Power Supply
Batteries 8 ea. 1.5 V mignon cell (8 ea. size AA)
(alkaline manganese per IEC LR14)
or 8 rechargeable NiMH batteries (must be
recharged externally)
Z502R charger Broad-range charger with barrel connec-
tor, input: 100 to 240 V AC,
output: 16.5 V DC, 1 A (Mascot)
Nominal range of use 8.5 to 12 V
Battery test Battery capacity display with battery sym-
bol in 4 segments: .
Querying of momentary battery voltage via
menu function.
Battery saver circuit Automatic shutdown of display illumination
after 10 to 30 seconds (after the last time
the rotary switch is actuated) can be
selected in the setup menu (see page 8).
The test instrument is automatically
switched to the standby mode if the mea-
sured value remains unchanged for
approximately 15 minutes, and if none of
the controls are activated during this time.
The instrument is switched off automatically
if the measured value remains unchanged
for a long period of time and if none of the
keys or the rotary switch have been acti-
vated during on-time (specified in sec-
onds).
Service life For R
INS
(1000 V / 1 MΩ) and R
LO
with
25 s on-time and 1 subsequent measure-
ment each for a duration of 5 s
– With batteries (alkaline manganese):
700 measurements
– With rechargeable batteries (2000 mAh):
650 measurements
Safety shutdown If supply voltage is too low (U < 8 V), the
instrument is switched off, or cannot be
switched on.
Recharging socket Installed rechargeable batteries can be
recharged directly by connecting a charger
to the recharging socket:
charger Z502R
Charging time Approx. 2 hours *
* Maximum charging time with fully depleted rechargeable batteries.
A timer in the charger limits charging time to no more than 4 hours.
Meas.
Qty.
U
iSO
Range Measuring Range
Reso-
lution
Open-Circuit
Voltage
U
0max
Test Current Intrinsic Uncertainty Measuring Uncertainty Overload Capacity
R
ISO
50 V
100 V
250 V / 500 V
1000 V
100 k 10 kΩ ... 99.9 kΩ 0.1 k
50 V /100 V:
1.25 U
ISO
250 V /
500 V /
1000 V:
1.1 U
ISO
I
N
= 1 mA
I
K
≤ 5 mA
±(5% rdg. + 3 d) ±(7% rdg. + 3 d)
1000 V AC/DC
TRMS
1 M 100 kΩ ... 999 kΩ 1 k
10 M 1.00 MΩ ... 9.99 MΩ 10 k
100 M 10.0 MΩ ... 99,9 MΩ 100 k
1 G 100 MΩ ... 999 MΩ 1 M
10 G 1.00 GΩ ... 9.99 GΩ 10 M
100 G 10,0 GΩ ... 99.9 GΩ 100 M ±(8% rdg. + 3 d)
1)
±(10% rdg. + 3 d)
1)
1 T 100 GΩ ... 999 GΩ 1 G ±(25% rdg. + 5 d)
1)
±
(50% rdg. + 20 d)
1) 2)
U
AC/DC
100 V 10.0 V ... 99.9 V 0.1 V
——±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
1000 V AC/DC
TRMS
3)
1000 V 100 V ... 999 V 1 V
R
LO
10 Ω 0.17 ... 9.99 Ω
0.01
Ω
4 V < U
0
< 6 V
200 mA ≤ I
I
≤
260 mA
4)
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
1000 V AC/DC
TRMS
R
Display range as of
01.0 Ω
100 Ω 10.0 ... 99.9 Ω 0.1 Ω
U
0
max. 15 V
1 mA ≤ I
I ≤ 1.3 mA
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
1000 V AC/DC
TRMS
1 kΩ 100 ... 999 Ω 1 Ω
10 kΩ 1.00 ... 9.99 kΩ 10 Ω
Parameter Range Intrinsic
Uncertainty
Measuring
Uncertainty
Voltage range 100 … 1000 V
±
(
10% rdg. + 8 d)
±
(
15% rdg. + 10 d)
Rise time 5 … 300 s — —
Measuring time 1 … 120 s /
a
uto / cont. measurement
——
t1 [min.] t2 [min.] Limit [min.]
PI 01:00 10:00 > 4.0 / > 3.0 / > 2.0 / > 1.5 / > 1.1 / > 1.0
DAR
00:30 01:00 > 1.60 / > 1.25