User Manual

GMC-I Messtechnik GmbH 27
1)
at 0 °C ... + 40 °C
2)
at input > 3.5 Vrms
3)
for < 10 kHz at 5 Vp-p
4)
without sensor
5)
display with current transformers 1000:1
Meas.
Func-
tion
Measuring
Range
DM61 DM62
(TRMS)
Resolu-
tion
Input
Impedance
Digital dis-
play inherent
deviation at
reference
condition
+(...% rdg.
+...digits)
Overload
Capacity
1)
Over-
load
Values
Over-
load
Dura-
tion
660.0 Ω 100 mΩ
–3.3 V 0.8 + 5
1000 V
DC
AC
eff/rms
sine
wave
max.
10 s
DIODE 2.000 V 1 mV 3.3 V 2.0 + 10
F
6.600 nF 1 pF
3.0 + 40
66.00 nF 10 pF 2.0 + 10
660.0 nF 100 pF 2.0 + 10
6.600 μF 1 nF 2.0 + 10
66.00 μF 10 nF 2.0 + 10
660.0 μF 100 nF 5.0 + 10
6.600 mF 1 μF 5.0 + 10
40.00 mF 10 μF 5.0 + 10
Hz
66.00 Hz 0.01 Hz
10 Hz (f min) 0.2 + 2
2)
660.0 Hz 0.1 Hz
6.600 kHz 1 Hz
66.00 kHz 10 Hz
660.0 kHz 100 Hz
6.600 MHz 1 kHz
10.00 MHz 10 kHz
%
1.0 ... 98.90%
0.01 %
0.9% (% min)
10 Hz ... 1 kHz
±5 digit
3)
1...10 kHz;
±
5 digit/kHz
°C/°F
0 ... 1300
°
C
••1°
C
—2,0
±
3
4)
–50 ... 0
°
C
••1°
C
—2,0
±
10
4)