User Manual

26 GMC-I Messtechnik GmbH
16 Technical Characteristics
Meas.
Func-
tion
Measuring
Range
DM61 DM62
(TRMS)
Resolu-
tion
Input
Impedance
Digital dis-
play inherent
deviation at
reference
condition
+(...% rdg.
+...digits)
Overload
Capacity
1)
Over-
load
Values
Over-
load
Dura-
tion
V(DC)
660.0 mV 100 μV
>100 M
Ω
//
<40pF
0.7 + 5
1000 V
DC
AC
eff/rms
sine
wave
cont.
6.600 V 1 mV
11 M
Ω
//<40pF
0.4 + 5
66.00 V 10 mV
10 M
Ω
// <40pF
0.4 + 5
660.0 V 100 mV
10 M
Ω
// <40pF
0.4 + 5
1000 V 1 V
10 M
Ω
// <40pF
0.4 + 5
V(AC)
660.0 mV 100 μV
>100 M
Ω
//
<40pF
1.2 + 5
6.600 V 1 mV
11 M
Ω
//<40pF
1.0 + 3
66.00 V 10 mV
10 M
Ω
// <40pF
660.0 V 100 mV
10 M
Ω
// <40pF
1000 V 1 V
10 M
Ω
// <40pF
A(DC)
Voltage Drop
66.00 mA 10 μA 66.00 mV 0.8 + 5
0.7 A cont.
660.0 mA 100 μA 66.00 mV 0.8 + 5
10.00 A 10 mA 10.00 mV 1.5 + 5 12 A cont.
A(AC)
66.00 mA 10 μA 66.00 mV 0.8 + 5
0.7 A cont.
660.0 mA 100 μA 66.00 mV 0.8 + 5
10.00 A 10 mA 10.00 mV 1.5 + 5 12 A cont.
(AC)
5)
66.00 A 10 mA 66.00 mV 0.8 + 5
0.7 A cont.
660.0 A 100 mA 66.00 mV 0.8 + 5
Ω
no-load Voltage
660.0 Ω 100 mΩ –3.3 V 0.8 + 5
1000 V
DC
AC
eff/rms
sine
wave
max.
10 s
6.600 kΩ ••1 Ω –1.08 V 0.8 + 5
66.00 kΩ 10 Ω –1.08 V 0.8 + 5
660.0 kΩ 100 Ω –1.08 V 0.8 + 5
6.600 MΩ ••1 kΩ –1.08 V 1.0 + 5
66.00 MΩ 10 kΩ –1.08 V 2.0 + 5