Datasheet

GMC-I Messtechnik GmbH 3
METRALINE DM 61/62
Analog-Digital Multimeter
Digital
Height of Main
Display numerals 7 segment numerals: 12 mm
Height of Sub
Display numerals 7 segment numerals: 7 mm
Number of counts 4 digit: 6600 steps
Overrange display „OL“ is shown
Polarity display “ sign is shown, When positive pole
connected to
Sampling rate 2.8 times/s
Power supply
Battery 2 AA size batteries alkaline manganese
cells as per IEC LR6.
Service life for METRALINE DM 61:
600 hrs. for V DC, A DC
300 hrs. for V AC, A AC
for METRALINE DM 62:
400 hrs. for V DC, A DC
200 hrs. for V AC, A AC
Battery test Automatic display of „ “ symbol when
battery voltage falls below following value:
approx. 2.4 V.
Electromagnetic compatibility (EMC)
Emission EN 61326: 2013 Class B
Immunity IEC 61000-4-2:
8 kV atmosphere discharge
4 kV contact discharge
IEC 61000-4-3: 3 V/m
Short-term measured value deviation may occur during electromagnetic interference
thus reducing the specified operating quality.
Safety: IEC 61010-1-2010
Measuring category 600 V CAT III, 300 V CAT IV
The maximum voltage of 1000 V may only
be used with CAT II.
High Voltage Test 6.7 kV (IEC 61010-1-2010)
Fuses
Fuse for up to 660 mA ranges
FF (UR) 1.6 A/1000 V AC/DC; 6.3 mm X 32 mm; rating 10 kA
with 1000 VAC/DC and ohmic load; in conjunction with power
diodes, protects all current measuring ranges up to 660 mA.
Fuse for up to 10 A ranges (METRALINE DM 62)
FF (UR)10 A/1000 V AC/DC; 10 mm x 38 mm; rating 30 kA with
1000 VAC/DC and ohmic load; protects the 10 A ranges up to
1000 V AC/DC.
Defective fuses are not displayed.
Response Time (after manual range selection)
Reference conditions
Ambient temperature 23 °C+ 2 K
Relative humidity 45% ... 55 % RH
Frequency of
measured quantity 50 or 60 Hz ±2%
Waveform of the
measured quantity sinusoidal
Battery voltage 3 V ±0.1 V
Environmental conditions
Functional
temperature range 0 °C ... +50 °C
Storage
temperature range –25 °C ... +70 °C (without batteries)
Relative humidity 45 ... 75 %
Altitude up to 2000 m
Mechanical configuration
Protection
for the meter IP50
Pollution degree 2
Connection sockets IP20 according to
EN 60529 / DIN VDE 0470-1
Dimensions with holster: 86 mm x 188 mm x 53 mm
without holster: 79 mm x 174 mm x 38 mm
Weight
480 g approx., including battery and holster
Applicable Regulations and Standards
Measured Quantity/
Measuring Range
Response Time
Transient response for
step function of the
measured quantity
Analog
Display
Digital Display
V , V , °C 0.1 s 1 s from 0 to 80%
of the upper range limit
A , A 0.1 s 1 s
from 0 to 50%
of the upper range limit
660 Ω ... 6.6 MΩ 0.1 s 1 s
66 MΩ 0.2 s 2 s
0.1 s 1 s
from 0 to 80%
of the upper range limit
6.6 nF ... 66 μF 0.7 s max. 1 s
660 μF ... 6.6 mF 1.4 s max. 3 s
66 mF 7.0 s max. 15 s
660 Hz, 6.6 kHz 2.0 s max. 2 s
66 kHz, 660 kHz, 1 MHz 0.5 s max. 1 s
% ( 10 Hz) 0.7 s max. 2.5 s
IEC 61010-1/EN 61010-1/
VDE 0411-1
Safety requirements for electrical equipment for
measurement, control and laboratory use
EN 60529
VDE 0470, Part 1
Test instruments and test procedures
Protection provided by enclosures (IP code)
DIN EN 61326-2-1
VDE 0843-02-2-1
Electrical equipment for measurement, control and labo-
ratory use – EMC requirements –
Part 2-1: Particular requirements for sensitive test and
measurement equipment
DIN EN 60529
DIN VDE 0470 Part 1
Test Instruments and test procedures
– Degree of protection provided by enclosures (IP code)