User Manual
26  GMC-I Messtechnik GmbH
16 Technical Characteristics   
Meas. 
Func-
tion
Measuring 
Range
DM61 DM62
(TRMS)
Resolu-
tion
Input
Impedance
Digital dis-
play inherent 
deviation at 
reference 
condition
+(...% rdg. 
+...digits)
Overload 
Capacity 
1)
Over-
load 
Values
Over-
load 
Dura-
tion
V(DC)
660.0 mV • • 100 μV
>100 M
Ω
 //
<40pF
0.7 + 5
1000 V
DC 
AC 
eff/rms
sine 
wave
cont.
6.600 V • • 1 mV
11 M
Ω
 //<40pF
0.4 + 5
66.00 V • • 10 mV
10 M
Ω
 // <40pF
0.4 + 5
660.0 V • • 100 mV
10 M
Ω
 // <40pF
0.4 + 5
1000 V • • 1 V
10 M
Ω
 // <40pF
0.4 + 5
V(AC)
660.0 mV • • 100 μV
>100 M
Ω
 //
<40pF
1.2 + 5
6.600 V • • 1 mV
11 M
Ω
 //<40pF
1.0 + 3
66.00 V • • 10 mV
10 M
Ω
 // <40pF
660.0 V • • 100 mV
10 M
Ω
 // <40pF
1000 V • • 1 V
10 M
Ω
 // <40pF
A(DC)
Voltage Drop
66.00 mA • • 10 μA 66.00 mV 0.8 + 5
0.7 A cont.
660.0 mA • • 100 μA 66.00 mV 0.8 + 5
10.00 A — • 10 mA 10.00 mV 1.5 + 5 12 A cont.
A(AC)
66.00 mA • • 10 μA 66.00 mV 0.8 + 5
0.7 A cont.
660.0 mA • • 100 μA 66.00 mV 0.8 + 5
10.00 A — • 10 mA 10.00 mV 1.5 + 5 12 A cont.
(AC) 
5)
66.00 A • — 10 mA 66.00 mV 0.8 + 5
0.7 A cont.
660.0 A • — 100 mA 66.00 mV 0.8 + 5
Ω
no-load Voltage
660.0 Ω • • 100 mΩ –3.3 V 0.8 + 5
1000 V
DC 
AC 
eff/rms
sine 
wave
max. 
10 s
6.600 kΩ ••1 Ω –1.08 V 0.8 + 5
66.00 kΩ • • 10 Ω –1.08 V 0.8 + 5
660.0 kΩ • • 100 Ω –1.08 V 0.8 + 5
6.600 MΩ ••1 kΩ –1.08 V 1.0 + 5
66.00 MΩ • • 10 kΩ –1.08 V 2.0 + 5