User Manual
GMC-I Messtechnik GmbH  17
The cut-out capacity of the fuse is 10 kA at a rated voltage 
of 1000 V AC/DC and ohmic load.
• The 10 A current measuring ranges are protected by a 
10 A / 1000 V AC/DC fuse. The cut-out capacity of the 
fuse is 30 kA at a nominal voltage of 1000 V AC/DC and 
ohmic load.
• Replacement of the fuses is described in section 17.
10.1 AC Current Measurement with (Clip-on) Current Transformer
10.1.1 Transformer Output mA/A
Caution
If current transformers are operated with an open circuit on the 
secondary side, e. g. due to defective or disconnected leads, a 
blown fuse in the multimeter, or a wrong connection, danger-
ously high voltages may occur at the connectors. Therefore, 
make sure that the current circuit of the multimeter and sec-
ondary winding of the transformer connected to the multimeter 
form an intact circuit. Connect the transformer to the   and 
mA or A sockets. The maximum permissible operating voltage 
is the nominal voltage of the current transformer. When reading 
the measured value, take into account the transformer ratio 
and the additional error in indication.
Transformer output METRALINE DM 61
The METRALINE DM 61 shows the switching position and the 
corresponding sockets. Connect a (clip-on) current trans-
former with a transmission ratio of 1000:1 to this socket. The 
measured values are then displayed directly in the „A“ range.