15.2 Creating and Transferring Test Sequences with
IZYTRONIQ (step-by-step instructions)
Select “Stationary Objects” .
Then select the “Sequences” menu .
Click the “Add” symbol . The “Create New Sequence” field
is displayed. Enter a “Sequence Name”, a “Test Type” and a
“Standard”, and select your currently connected instrument
under “For Device”. Acknowledge by clicking “Add”.
Save the settings by clicking the icon.
Select the new entry and then the sequence editor . The
editing menu appears with “Step Selection” and “Design
Progress”.
Select the test instrument which is displayed in “Step Selec-
tion”. “Visual Inspection” and “User-Evaluated Measurement”
appear.
“Test Step: Visual Inspection” is opened in the bottom left-
hand window by dragging “Visual Inspection” into the “Design
Progress” field. The parameters or details for the respective
test step must be entered here.
Save the settings by clicking the icon.
“Test Step: User-Evaluated Measurement” is opened in the
bottom left-hand window by dragging “User-Evaluated Mea-
surement” into the “Design Progress” field. The parameters or
details for the respective test step must be entered here.
Save your settings.
Repeat the test steps until the test sequence is finished.
Save your settings by clicking the icon.
Select “Stationary Objects” again .
Then select the “Export” function . The export wizard ap-
pears.
Select the desired test instrument and insert a checkmark
next to “Sequences”. Select “Export. The “Export Sequences
(max. 10)” menu appears.
Mark the sequences to be exported and click the “Export to
Test Instrument” icon .
During the transfer of the
test sequences a prog-
ress bargraph is shown
at the PC screen and
the right-hand image
appears on the display
of the test instrument.
Successful transfer to
the test instrument via
IZYTRONIQ is then indi-
cated at the PC.
Note
All test sequences previously stored at the test instru-
ment are deleted. Only those test sequences transferred
during the most recent export operation via IZYTRONIQ are
saved at the test instrument.