User Manual
14 GMC-I Messtechnik GmbH
5.5 Help Function
The following information can be displayed for each switch posi-
tion and basic function after it has been selected with the rotary selec-
tor switch:
• Wiring diagram
• Measuring range
• Nominal range of use and measuring uncertainty
• Nominal value
➭ Press the HELP key in order to query online help:
➭ If several pages of help are available for the respective mea-
suring function, the HELP key must be pressed repeatedly.
➭ Press the ESC key in order to exit online help.
5.6 Setting Parameters or Limit Values using RCD Measurement as an Example
1 Access the submenu for setting the desired parameter.
2 Select a parameter using the ↑ or ↓ scroll key.
3 Switch to the setting menu for the selected parameter with the → scroll
key.
4 Select a setting value using the ↑ or ↓ scroll key.
5 Acknowledge the setting value with the ↵ key. This value is transferred to
the setting menu.
6 The setting value is not permanently accepted for the respective measure-
ment until
✓ is pressed, after which the display is returned to the main
menu. You can return to the main menu by pressing ESC instead of
✓,
without accepting the newly selected value.
Parameter Lock (plausibility check)
Individually selected parameter settings are checked for plausibil-
ity before transfer to the measurement window.
If you select a parameter setting which doesn’t make sense in
combination with other parameter settings which have already
been entered, it’s not accepted. The previously selected parame-
ter setting remains unchanged.
Remedy: Select another parameter setting.
1
2
2
3
4
4
5
6
2
4
3
5
6