Please note that the test sequences loaded to the test instrument
are deleted by the following operations in the test instrument:
• by receiving new test sequences from the PC
• by resetting to default settings
(Switch position SETUP → key GOME SETTING)
• by firmware updates
• by changing the user language
(Swith position SETUP → key CULTURE)
• by deleting the entire database in the test instrument
Test Sequence Commands
Figure 15.3
Confirm note
Discard an event
Confirm event
Move to previous/
next step
Save measurement
result
Configuring Test Sequence Parameters
Measurement parameters are also configured at the PC. How-
ever, parameters can be changed at the test instrument during
the test sequence before the respective measurement is started.
When the test step is started once more, the parameter settings
specified in IZYTRONIQ are loaded again.
Note
IZYTRONIQ does not subject the parameters to a plausibil-
ity check. As a result, the newly created test sequence
should be checked at the test instrument before it’s per-
manently added to the database.
Limit values are not currently set in IZYTRONIQ, and have to be
adjusted during the automatic test sequence.
Selecting and Starting a Test Sequence at the Test Instrument
Figure 15.4
Press the START key to launch the selected test sequence (here:
SEQU.1).
When executing a test step of the measurement type, the display
structure known from the individual measurements is shown.
Instead of the storage and battery symbol, the current test step
number is shown in the header (here: step 01 of 06), see figure
15.2. The next test step is shown after pressing the „Save“ key
twice.
Setting Parameters and Limit Values
Parameters and limit values can also be modified while performing
a test sequence or before starting the measurement. This modifi-
cation only affects the active test procedure and is not saved.
Skipping of Test Steps
There are two possibilities to skip test steps and/or individual
measurements:
• Activate test sequence, switch to the right-hand test step col-
umn with the cursor, select the x
th
test step and press key
START.
• Within a test sequence, the navi-
gation menu is activated by press-
ing the navigation key Cursor left-
right. Switch to the previous or
next test step with the cursors
which are now displayed sepa-
rately.
Leave the navigation menu and reactivate the current test step
with ESC.
Interrupt or Abort a Test Sequence
An active sequence is aborted with ESC and subsequent confir-
mation.
When the last test step is completed, the message „Sequence
completed“ is shown. After confirming this message, the start
menu „List of test sequences“ appears on the display.
Figure 15.5