15.2 Creating and Transferring Test Sequences with
IZYTRONIQ (step-by-step instructions)
➭ Select “Stationary Objects” .
➭ Then select the “Sequences” menu .
➭ Click the “Add” symbol . The “Create New Sequence” field
is displayed. Enter a “Sequence Name”, a “Test Type” and a
“Standard”, and select your currently connected instrument
under “For Device”. Acknowledge by clicking “Add”.
➭ Save the settings by clicking the icon.
➭ Select the new entry and then the sequence editor . The
editing menu appears with “Step Selection” and “Design
Progress”.
➭ Select the test instrument which is displayed in “Step Selec-
tion”. “Visual Inspection” and “User-Evaluated Measurement”
appear.
➭ “Test Step: Visual Inspection” is opened in the bottom left-
hand window by dragging “Visual Inspection” into the “Design
Progress” field. The parameters or details for the respective
test step must be entered here.
➭ Save the settings by clicking the icon.
➭ “Test Step: User-Evaluated Measurement” is opened in the
bottom left-hand window by dragging “User-Evaluated Mea-
surement” into the “Design Progress” field. The parameters or
details for the respective test step must be entered here.
➭ Save your settings.
➭ Repeat the test steps until the test sequence is finished.
➭ Save your settings by clicking the icon.
➭ Select “Stationary Objects” again .
➭ Then select the “Export” function . The export wizard ap-
pears.
➭ Select the desired test instrument and insert a checkmark
next to “Sequences”. Select “Export”. The “Export Sequences
(max. 10)” menu appears.
➭ Mark the sequences to be exported and click the “Export to
Test Instrument” icon .
During the transfer of the
test sequences a prog-
ress bargraph is shown
at the PC screen and
the right-hand image
appears on the display
of the test instrument.
Successful transfer to
the test instrument via
IZYTRONIQ is then indi-
cated at the PC.
Note
All test sequences previously stored at the test instru-
ment are deleted. Only those test sequences transferred
during the most recent export operation via IZYTRONIQ are
saved at the test instrument.