5.7 Freely Selectable Parameter Settings or Limit Values
5.7.1 Changing Existing Parameters
Individual parameters can be changed for certain measuring func-
tions, i.e. adjusted within predetermined limits.
The EDIT menu doesn’t appear until after switching to the
right-hand column and selecting the editable parameter .
Example for RLO Measuring Function – Parameter: LIMIT RLO
1 Open the submenu for setting the desired parameter (no fig-
ure, see section 5.6).
2 Select the editable parameter (identified with the icon)
with the or scroll key.
3 Select the edit menu by pressing the key.
4 Select the respective characters with the left or right cursor
key. The character is accepted by pressing the key. The
value is acknowledged by selecting
and then pressing the
key.
Note
Observe the predefined limits for the new setting value.
Enter any places to the right of the decimal point as well.
5.7.2 Adding New Parameters
For certain measuring functions, additional values within pre-
defined limits can be added in addition to the fixed values.
The EDIT menu doesn’t appear until after switching to the
right-hand column.
Example for Measuring Function I
ΔN
– Parameter: I
ΔN
1 Open the submenu for setting the desired parameter (no fig-
ure, see section 5.6).
2 Select the edit menu by pressing the key.
3 Select the respective characters with the left or right cursor
key. The character is accepted by pressing the key. The
value is acknowledged by selecting
and then pressing the
key. The new parameter is added to the list.
Note
Observe the predefined limits for the new setting value.
Enter any places to the right of the decimal point as well.
Select the EDIT menu.
Select value / U/M.
Select value / U/M.
Accept value / U/M.
Delete characters.
Save value (to list).
Select the EDIT+ menu.
Select value / U/M.
Select value / U/M.
Accept value / U/M.
Delete characters.
Save value (to list).