5.5 Help Function
The following information can be displayed for each switch posi-
tion and basic function after it has been selected with the rotary selec-
tor switch:
Wiring diagram
Measuring range
Nominal range of use and measuring uncertainty
Nominal value
Press the HELP key in order to query online help:
If several pages of help are available for the respective mea-
suring function, the HELP key must be pressed repeatedly.
Press the ESC key in order to exit online help.
5.6 Setting Parameters or Limit Values using RCD Measurement as an Example
1 Access the submenu for setting the desired parameter.
2 Select a parameter using the orscroll key.
3 Switch to the setting menu for the selected parameter with the scroll
key.
4 Select a setting value using the orscroll key.
5 Acknowledge the setting value with the key. This value is transferred to
the setting menu.
6 The setting value is not permanently accepted for the respective measure-
ment until
is pressed, after which the display is returned to the main
menu. You can return to the main menu by pressing ESC instead of
,
without accepting the newly selected value.
Parameter Lock (plausibility check)
Individually selected parameter settings are checked for plausibil-
ity before transfer to the measurement window.
If you select a parameter setting which doesn’t make sense in
combination with other parameter settings which have already
been entered, it’s not accepted. The previously selected parame-
ter setting remains unchanged.
Remedy: Select another parameter setting.