Data Sheet

PROFITEST MTECH+, MPRO, MXTRA, SECULIFE IP
DIN VDE 0100/IEC 60364-6 Testers
8 GMC-I Messtechnik GmbH
PROFITEST MASTER Characteristic Values
Reference Conditions
Line voltage 230 V ± 0.1 %
Line frequency 50 Hz ± 0.1 %
Meas. quantity frequency
45 Hz 65 Hz
Measured qty. waveform
Sine (deviation between effective and
rectified value 0.1 %)
Line impedance angle cos ϕ =1
Probe resistance 10 Ω
Supply power 12 V ± 0.5 V
Ambient temperature + 23° C ± 2 K
Relative humidity 40% to 60%
Finger contact For testing potential difference
to ground potential
Standing surface
insulation Purely ohmic
Power Supply
Rechargeable batteries 8 each AA 1.5 V,
we recommend only using the battery
pack included in the standard equip-
ment (pack of rechargeable batteries
article no. Z502H)
Number of measurements (standard setup with illumination)
For R
ISO
1 measurement – 25 s pause:
Approx. 1100 measurements
– For R
LO
Automatic polarity reversal / 1 Ω
(1 measuring cycle) – 25 s pause:
Approx. 1000 measurements
Battery test Symbolic display of battery voltage
Battery saver circuit
Display illumination can be switched off.
The test instrument is switched off
automatically after the last key opera-
tion. The user can select the desired
on-time.
Safety shutdown If supply voltage is too low, the instru-
ment is switched off, or cannot be
switched on.
Recharging socket Installed rechargeable batteries can be
recharged directly by connecting a
charger to the recharging socket:
charger Z502R
Charging time Charger Z502R:
Approx. 2 hours *
* Maximum charging time with fully depleted rechargeable batteries.
A timer in the charger limits charging time to no more than 4 hours.
Overload Capacity
R
ISO
1200 V continuous
U
L-PE
, U
L-N
600 V continuous
RCD, R
E
, R
F
440 V continuous
Z
L-PE
, Z
L-N
550 V (Limits the number of measure-
ments and pause duration. If overload
occurs, the instrument is switched off
by means of a thermostatic switch.)
R
LO
Electronic protection prevents switching
on if interference voltage is present
Fine-wire
fuse protection FF 3.15 A 10 s, fuses blow at > 5 A
Electrical Safety
Protection class II per IEC 61010-1/EN 61010-1/
VDE 0411-1
Nominal voltage 230/400 V (300/500 V)
Test voltage 3.7 kV 50 Hz
Measuring category CAT III 500 V or CAT IV 300 V
Pollution degree 2
Fusing, L and N terminals
1 cartridge fuse-link ea.
FF 3.15/500G 6.3 x 32 mm
Electromagnetic Compatibility (EMC)
Product standard EN 61326-1:2013
Ambient Conditions
Accuracy 0 to + 40 °C
Operation –5 to + 50 °C
Storage 20 to +60 °C (without rechargeable
batteries)
Relative humidity
Max. 75%, no condensation allowed
Elevation Max. 2000 m
Mechanical Design
Display Multiple display with dot matrix,
128 x 128 pixels
Dimensions W x L x D: 260 x 330 x 90 mm
Weight approx. 2.7 kg
with rechargeable batteries
Protection Housing: IP 40, test probe: IP 40 per
EN 60529/DIN VDE 0470, part 1
Data Interfaces
Type USB slave for PC connection
Type RS 232 for barcode and RFID scanners
Type Bluetooth
®
for connection to PC
(PROFITEST MTECH+/MXTRA/SECULIFE IP
only)
BAT
Interference emission Class
EN 55022 A
Interference immunity Test Value Feature
EN 61000-4-2 Contact/atmos. – 4 kV/8 kV
EN 61000-4-3 10 V/m
EN 61000-4-4 Mains connection – 2 kV
EN 61000-4-5 Mains connection – 1 kV
EN 61000-4-6 Mains connection – 3 V
EN 61000-4-11 0.5 period / 100%